DOI 10.1097/mao.0000000000002247
Temporal Bone Histopathology of First-Generation Cochlear Implant Electrode Translocation.
Аннотация
OBJECTIVE
To evaluate the histopathology of human temporal bones (HTBs) with cochlear implants (CI).
BACKGROUND
Understanding CI translocation injuries is critical for improving outcomes.
MATERIAL AND
METHODS
Thirteen HTBs from 12 CI patients were studied. Six HTBs exhibited translocation with localized injury (Group 1) and seven HTBs exhibited translocation with significant lateral wall injury (Group 2). There were no significant differences between Group 1 and Group 2 for age at death, age at implantation, and years with CI.
RESULTS
Four out of six of Group 1 had round window approach, while all seven of Group 2 had cochleostomy approach. Translocation injuries tended to occur near 180 degrees of angular insertion with a mean of 186.36 ± 51.62 degrees. Average CI insertion length for Group 2 was 21.86 ± 2.55 mm, significantly longer than Group 1 at 18.50 ± 3.33 mm (p = 0.031). Group 1 had an average of 17300 ± 9415 spiral ganglia neurons (SGNs) while Group 2 had significantly fewer SGNs 6714 ± 4269 (p = 0.015). Group 1 average auditory performance of 66.55 ± 27.20% was higher than that of Group 2 of 39.86 ± 15.36%. Group 2 had a high degree of osteoneogenesis and infiltration of cells generally localized to areas of translocation injury and cochleostomy.
CONCLUSION
Translocation injuries tend to occur at an insertion angle of 180 degrees, at 9 to 10 mm. Lateral wall injury and damage to the organ of Corti incites fibrosis, osteoneogenesis, and infiltration, lower SGN count and poorer auditory performance. Longer electrodes were more prone to translocation and higher chance of significant intracochlear injury.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.