Differences in the Effectiveness of Smoke Evacuator Types for Reducing Surgical Smoke: A Randomized Controlled Trial.

Yoshizaki Yuhi, Kawaguchi Yoshikuni, Kawakami Toshiaki, Hayakawa Tomoaki, Hayashi Yoshihito, Sawa Yui, Ito Kyoji, Kashiwabara Kosuke, Nishikata Yasuyoshi, Hasegawa Kiyoshi

СтатьяРандомизированное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Surgical smoke contains respirable particles and volatile organic compounds (VOCs) associated with carcinogenic and infectious risks to operating room personnel. Whether electrocautery-incorporated smoke evacuation improves particulate and VOC clearance compared with hose-type evacuation remains undefined.

STUDY

DESIGN

In this randomized, double-blind clinical trial (UMIN000047033) conducted from February 2022 to May 2023 at The University of Tokyo Hospital, 64 patients undergoing laparotomy were assigned 1:1:1 to electrocautery-incorporated evacuation with an 8 mm tip distance (n = 22), 18 mm tip distance (n = 23), or hose-type evacuation (n = 21). The primary endpoint was the change in particle counts at 0.3, 0.5, 1.0, and 5.0 μm during a standardized 10-minute sampling period. Secondary endpoints were acetaldehyde and formaldehyde levels. Comparisons used the Wilcoxon rank-sum testing with Bonferroni correction.

RESULTS

The 8 mm electrocautery-incorporated system reduced particle counts by 96% to 97% across 0.3 to 5.0 μm compared with hose-type evacuation (all p < 0.05). The 18 mm system reduced 0.3 μm particles by 78% vs hose (p = 0.005), with 81% to 88% reductions for 0.5 to 5.0 μm particles (p > 0.05). Median (interquartile range) acetaldehyde levels were 1.90 (1.10 to 2.50), 2.70 (0.50 to 4.00), and 7.90 (3.10 to 15.20) μg/m3 in the 8 mm, 18 mm, and hose groups, respectively; both electrocautery groups were lower than hose (all p < 0.05). Formaldehyde levels were similarly reduced (all p < 0.05).

CONCLUSIONS

Electrocautery-incorporated smoke evacuation significantly reduced respirable particle and VOC exposure compared with hose-type evacuation, with maximal particulate reduction at an 8 mm tip distance. Device selection and proximity to the cautery source materially influence operating room exposure to surgical smoke.

Для цитирования: Yoshizaki Yuhi, Kawaguchi Yoshikuni, Kawakami Toshiaki, Hayakawa Tomoaki, Hayashi Yoshihito, Sawa Yui, Ito Kyoji, Kashiwabara Kosuke, Nishikata Yasuyoshi, Hasegawa Kiyoshi. Differences in the Effectiveness of Smoke Evacuator Types for Reducing Surgical Smoke: A Randomized Controlled Trial.. Journal of the American College of Surgeons. 2026. DOI: 10.1097/xcs.0000000000001943.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.