Evaluating surgical smoke in laser-assisted laparoscopic partial nephrectomy with a novel handle: anin Vitrostudy.
Аннотация
OBJECTIVES
Lasers show promising application in partial nephrectomy due to their excellent hemostatic effect. This study aimed to evaluate the suction effect of the novel laser operating handle on the smoke generated by a 980-nm diode laser (DL) and a 532-nm potassium-titanyl-phosphate (KTP) green laser (GL) in laparoscopic partial nephrectomy (LPN) and to analyse the surgical smoke generated by the laser treatment.
METHODS
Fresh porcine kidneys were purchased from a slaughterhouse. Electrocoagulation knife (EK), ultrasonic knife (UK), DL, and GL were utilised in LPN. The operation time and the composition and concentration of the surgical smoke were recorded. HE staining was performed to observe the degree of thermal damage.
RESULTS
In terms of smoke generation, the 532-nm KTP laser results in the largest amount of smoke. The surgical field of DL-LPN was adequately improved with the aid of the instrument with the novel handle. The thickness of the coagulation layer, which ranged from thick to thin, was DL, GL, EK, and UK. The coagulation of EK, UK, DL, and GL was 0.35 ± 0.08 mm, 0.52 ± 0.08 mm, 4.16 ± 0.21 mm, and 1.55 ± 0.21 mm, respectively. For formaldehyde, total volatile organic compounds (TVOC), and carbon monoxide (CO), there was a marked reduction after the application of the aspirator. For the DL group, the Formaldehyde TVOC and CO were 4.95 ± 0.48 mg/m3 and 1.69 ± 0.46 mg/m3, 8.34 ± 0.72 mg/m3 and 1.11 ± 0.33 mg/m3, 124.14 ± 9.58 PPM and 23.71 ± 3.9 PPM, respectively.
CONCLUSIONS
As shown in the present study, though a large amount of harmful smoke was generated during laser-assisted laparoscopic partial nephrectomy (LALPN), the novel laser handle can be sucked effectively and provides a clear surgical field. As to the harmful gases released, an effective protective mask is needed.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.