A novel technique to measure in vivo uterine suspensory ligament stiffness.
Аннотация
OBJECTIVE
The purpose of this study was to describe a new computer-controlled research apparatus for measuring in vivo uterine ligament force-displacement behavior and stiffness and to present pilot data for women with and without prolapse.
STUDY
DESIGN
Seventeen women with varying uterine support underwent testing in the operating room (OR) after anesthetic induction. A tripod-mounted computer-controlled linear servoactuator was used to quantify force-displacement behavior of the cervix and supporting ligaments. The servoactuator applied a caudally directed force to a tenaculum at 4 mm/sec velocity until the traction force reached 17.8 N (4 lbs). Cervix location on Pelvic Organ Prolapse Quantification system (POP-Q) in the clinic, in the OR, at rest, and with minimal force (<1.1 N); maximum force (17.8 N) was recorded. Ligament "stiffness" between minimum and maximum force was calculated.
RESULTS
The mean ± SD subject age was 54.5 ± 12.7 years; parity was 2.9 ± 1.1; body mass index was 29.0 ± 4.3 kg/m(2), and POP-Q point C was -3.1 ± 3.9 cm. POP-Q point C was correlated most strongly with cervix location at maximum force (r = +0.68; P = .003) and at rest (r = +0.62; P = .009). Associations between cervix location at minimum force (r = +0.46; P = .059) and ligament stiffness (r = -0.44; P = .079) were not statistically significant. Cervix location in the OR with minimal traction lay below the lowest point found on POP-Q for 13 women.
CONCLUSION
POP-Q point C was correlated strongly with cervix location at rest and at maximum traction force; however, only 19% of the variation in POP-Q point C location was explained by ligament stiffness. The cervix location in the OR at minimal traction lay below POP-Q point C value in three-fourths of the women.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.