MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1117/1.jbo.30.7.076004

Медицинская библиотека · Journal of biomedical optics · 2025

Automatic photoacoustic monitoring of perinatal brain hypoxia with superior sagittal sinus detection.

Jiang Baichuan, Graham Ernest, Unberath Mathias, Taylor Russell H, Koehler Raymond C, Kang Jeeun, Boctor Emad M

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод в очереди

Аннотация

SIGNIFICANCE: Despite advances in perinatal medicine over decades, perinatal hypoxic-ischemic encephalopathy (HIE) remains a significant cause of fetal cerebral palsy and can lead to other severe medical sequelae or death. Therefore, it is highly desirable to effectively detect brain hypoxia during labor and postnatally for HIE management.

AIM: We recently validated the feasibility of transcranial photoacoustic (PA) imaging for oxyhemoglobin saturation measurement at the superior sagittal sinus ( O 2 Sat ss ) in the neonatal piglet brain, at which overall oxygen supply status can be reflected as a primary collective vein. We aim to automate the PA-based workflow of at-risk subject detection and enable fully autonomous and continuous perinatal monitoring.

APPROACH: We proposed a two-step algorithm that focuses on the most informative region of the brain for oxygenation status, the superior sagittal sinus (SSS). First, a convolutional neural network (U-Net) is trained to detect the location of SSS in the coronal cross-section PA images. Then, an optimized region of interest patch around the predicted SSS location is cropped from the spectral unmixed image and averaged as the O 2 Sat ss measurement. A confidence score can be computed for the measurement via Monte Carlo dropout (MCD), which infers the prediction uncertainty for better clinical decision-making.

RESULTS

The algorithm was evaluated on an in vivo piglet brain imaging dataset containing 84 independent experimental settings from 10 piglet subjects. A 10-fold leave-one-subject-out cross-validation experiment reports 85.2% sensitivity and 93.3% specificity for healthy/hypoxia classification with an R -squared value of 0.708 and a confidence score of 94.06% based on MCD computation, well agreed with our ground-truth given by blood gas measurements.

CONCLUSIONS

The proposed automatic O 2 Sat ss monitoring solution demonstrated a hypoxia detection capability comparable to the human expert manual annotation on the same task. We concluded with high feasibility for a noninvasive PA-based continuous monitoring of the perinatal brain.

Для цитирования: Jiang Baichuan, Graham Ernest, Unberath Mathias, Taylor Russell H, Koehler Raymond C, Kang Jeeun, Boctor Emad M. Automatic photoacoustic monitoring of perinatal brain hypoxia with superior sagittal sinus detection.. Journal of biomedical optics. 2025. DOI: 10.1117/1.jbo.30.7.076004.