DOI 10.1111/trf.15583

Медицинская библиотека · Transfusion · 2019

A dual-center evaluation of platelet culture vials to detect the presence of microorganisms in platelets.

Denys Gerald, Tulpule Anagha, Roth Jessica, Warns Patty, Collins Tiffany, Mindel Susan

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Microorganism contamination of platelets results in a high risk of transfusion-related sepsis. Here, the ability of culture vials (BD BACTEC Platelet Aerobic/F and Platelet Anaerobic/F vials, Becton, Dickinson and Company) to detect microorganisms in leukoreduced apheresis platelets (LRAPs) and leukoreduced whole blood platelet concentrates (LRWBPCs) was assessed.

METHODS

LRAPs or LRWBPCs were inoculated into Aerobic/F and Anaerobic/F vials and placed in a blood culturing system (BD BACTEC FX System, Becton, Dickinson and Company) for growth/monitoring over 7 days to detect preexisting contamination during false-positive testing. Subsequently, platelets were seeded with microorganisms at approximately 10 CFU/mL or approximately 1 CFU/mL to simulate contamination. Aerobic/F and Anaerobic/F vials were inoculated with platelets (sets of 12). Microorganism growth was detected in the BACTEC FX instrument over 7 days. Overall, 2925 vials were tested.

RESULTS

Of the 1905 vials included in the microorganism detection phase, 63 (3.3%) Aerobic/F and 16 (0.8%) Anaerobic/F vials were both BACTEC FX and subculture negative. From the remaining 1827 vials, two (0.1%) Anaerobic/F vials were false positive; no false positives were observed in Aerobic/F vials, and no false negatives occurred in either vial type. Of the remaining 1825 vials (99.9%), 955 Aerobic/F and 870 Anaerobic/F vials were true positives. The mean-time-to-detection range was 8.5 to 77 hours. All true-positive Aerobic/F and Anaerobic/F vials showed 100% agreement with subculture for positive identification of seeded microorganisms.

CONCLUSION

Aerobic/F and Anaerobic/F vials facilitate contamination detection in LRAPs and LRWBPCs down to approximately 1 CFU/mL. These results support the use of Aerobic/F and Anaerobic/F vials for quality control testing of platelets before transfusion.

Для цитирования: Denys Gerald, Tulpule Anagha, Roth Jessica, Warns Patty, Collins Tiffany, Mindel Susan. A dual-center evaluation of platelet culture vials to detect the presence of microorganisms in platelets.. Transfusion. 2019. DOI: 10.1111/trf.15583.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.