Technical and clinical performance of the thermo-test device "Q-Sense" to assess small fibre function: A head-to-head comparison with the "Thermal Sensory Analyzer" TSA in diabetic patients and healthy volunteers.

Pfau Doreen B, Greffrath Wolfgang, Schilder Andreas, Magerl Walter, Ohler Carolin, Westermann Andrea, Maier Christoph, Doppler Kathrin, Sommer Claudia, Orth Michael, Hammes Hans-Peter, Kurz Jochen, Götz Marcus, Treede Rolf-Detlef, Schuh-Hofer Sigrid

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Thermo-test devices are rarely used outside specialized pain centres because of high acquisition costs. Recently, a new, portable device ("Q-Sense") was introduced, which is less expensive but has reduced cooling capacity (20°C). We assessed the reliability/validity of the "Q-Sense" by comparing it with the Thermal Sensory Analyzer (TSA).

METHODS

Using a phantom-skin model, the physical characteristics of both devices were compared. The clinical performance was assessed in a multicentre study by performing Quantitative Sensory Testing (QST) in 121 healthy volunteers and 83 diabetic patients (Eudra-Med-No. CIV-12-05-006501).

RESULTS

Both device types showed ~40% slower temperature ramps for heating/cooling than nominal data. Cold/warm detection thresholds (CDT, WDT) and heat pain thresholds (HPT) of healthy subjects did not differ between device types. Cold pain thresholds (CPT) were biased for Q-Sense by a floor effect (p < .001). According to intraclass correlation coefficients (ICC), agreement between TSA and Q-Sense was good/excellent for CDT (ICC = 0.894) and WDT (ICC = 0.898), moderate for HPT (ICC = 0.525) and poor for CPT (ICC = 0.305). In diabetic patients, the sensitivity of Q-Sense to detect cold hypoesthesia was reduced in males >60 years. Moderate correlations between thermal detection thresholds and morphological data from skin biopsies (n = 51) were similar for both devices.

CONCLUSIONS

Physical characteristics of both thermo-test devices are similarly limited by the poor temperature conduction of the skin. The Q-Sense is useful for thermal detection thresholds but of limited use for pain thresholds. For full clinical use, the lower cut-off temperature should be set to ≤18°C.

SIGNIFICANCE: High purchase costs prevent a widespread use of thermo-test devices for diagnosing small fibre neuropathy. The air-cooled "Q-Sense" could be a lower cost alternative, but its technical/clinical performance needs to be assessed because of its restricted cut-off for cooling (20°C). This study provides critical information on the physical characteristics and the clinical validity/reliability of the Q-Sense compared to the "Thermal Sensory Analyzer" (TSA). We recommend lowering the cut-off value of the Q-Sense to ≤18°C for its full clinical use.

Для цитирования: Pfau Doreen B, Greffrath Wolfgang, Schilder Andreas, Magerl Walter, Ohler Carolin, Westermann Andrea, Maier Christoph, Doppler Kathrin, Sommer Claudia, Orth Michael, Hammes Hans-Peter, Kurz Jochen, Götz Marcus, Treede Rolf-Detlef, Schuh-Hofer Sigrid. Technical and clinical performance of the thermo-test device "Q-Sense" to assess small fibre function: A head-to-head comparison with the "Thermal Sensory Analyzer" TSA in diabetic patients and healthy volunteers.. European journal of pain (London, England). 2019. DOI: 10.1002/ejp.1461.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.