Biodegradable spacer reduces the subacromial pressure: A biomechanical cadaver study.

Chevalier Yan, Pietschmann Matthias F, Thorwächter Christoph, Chechik Ofir, Adar Eliyahu, Dekel Assaf, Müller Peter E

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Failure after rotator cuff repair remains a major clinical problem and could be related to excessive pressures from the acromion. Previous studies with irreparable tears showed good clinical results of tendon healing with arthroscopic insertion of a protective biodegradable spacer balloon between the repaired tendon and the acromion. One hypothesis is that compression pressures on the repaired tendon will be reduced by the spacer. This cadaver study aimed to investigate the effects of this subacromial spacer on compression pressures over a repaired supraspinatus tendon in passive motion.

METHODS

Rotator cuff tear and repair were performed in six fresh-frozen cadaveric shoulders, followed by insertion of a biodegradable subacromial spacer. Specimens were tested using a passive shoulder simulator for abduction-adduction, flexion-extension and internal-external rotations. A sensor positioned below the acromion was used to measure compression pressure changes through passive range of motion before and after placement of a subacromial spacer. Peak pressures were measured in adduction-abduction motion, near 90° abduction.

FINDINGS: Both the mean and peak pressures in abduction-adduction were significantly reduced after insertion of the subacromial spacer (from mean 121.7 (SD 9.5) MPa to 51.5 (SD 1.2) MPa and from peak 1749.6 (SD 80.7) MPa to 535.1 (SD 27.6) MPa) (P<0.0001).

INTERPRETATION: The reduced peak pressures and wider load distributions over the sensor during both passive abduction-adduction and flexion-extension motions suggest that the use of the spacer will lead to reduced wear of the repair in patients, and potentially prevent rotator cuff re-tear after surgical repair.

Для цитирования: Chevalier Yan, Pietschmann Matthias F, Thorwächter Christoph, Chechik Ofir, Adar Eliyahu, Dekel Assaf, Müller Peter E. Biodegradable spacer reduces the subacromial pressure: A biomechanical cadaver study.. Clinical biomechanics (Bristol, Avon). 2017. DOI: 10.1016/j.clinbiomech.2017.12.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.