Percutaneous management of RetroFlex 3 balloon rupture and separation of the Edwards Sapien delivery system.

Green Kelly D, Waked Alain, Majeed Usman, Fudim Marat, Robbins Mark A, Crenshaw Marshall, Zhao David

Клинический случайСтатьяVideo-Audio MediaПеревод в очереди

Аннотация

We report the case of an 85-year-old woman with severe aortic stenosis who underwent transcatheter aortic valve replacement with use of the Edwards Sapien(®) valve system. The procedure was complicated by rupture of the valve-deployment balloon, with separation and retention of the nose cone of the RetroFlex 3(®) delivery system in the iliac artery. Our endovascular retrieval of the equipment was successful, and we achieved access-site hemostasis by deploying a covered stent. To our knowledge, this is the first report of the endovascular retrieval of a malfunctioning delivery system during transcatheter aortic valve replacement.

Для цитирования: Green Kelly D, Waked Alain, Majeed Usman, Fudim Marat, Robbins Mark A, Crenshaw Marshall, Zhao David. Percutaneous management of RetroFlex 3 balloon rupture and separation of the Edwards Sapien delivery system.. Texas Heart Institute journal. 2014. DOI: 10.14503/thij-13-3447.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.