DOI 10.2196/60455

Медицинская библиотека · JMIR formative research · 2025

Remote Photoplethysmography Technology for Blood Pressure and Hemoglobin Level Assessment in the Preoperative Assessment Setting: Algorithm Development Study.

Tan Selene Y L, Chai Jia Xin, Choi Minwoo, Javaid Umair, Tan Brenda Pei Yi, Chow Belinda Si Ying, Abdullah Hairil Rizal

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Blood pressure (BP) and hemoglobin concentration measurements are essential components of preoperative anesthetic evaluation. Remote photoplethysmography (rPPG) is an emerging technology that may be used to measure BP and hemoglobin concentration noninvasively with just a consumer-grade smartphone, replacing traditional in-person measurements. However, there is limited data regarding the use of this technology in patients with diverse skin tones and medical comorbidities. Hence, widespread applicability is yet to be achieved. The potential benefits of achieving this would be immense, allowing for greater convenience, accessibility, and reduction in labor and resources.

OBJECTIVE

Our study aims to be the first to develop an algorithm for noninvasive rPPG-based BP and hemoglobin concentration measurement that can be used for preoperative evaluation of patients in real-world clinical practice settings.

METHODS

We conducted the study at Singapore General Hospital from March 1, 2023, to June 28, 2024. A total of 200 patients were recruited. Our primary analysis compared the accuracy of rPPG-based systolic and diastolic BP measurements against measurements taken with automated BP measuring devices. Our secondary analysis compared the accuracy of rPPG-based hemoglobin concentration measurement against traditional blood sampling.

RESULTS

Our model performed best with diastolic BP predictions, with a mean absolute percentage error of 7.52% and a mean difference of 0.16 mm Hg (SD 3.22 mm Hg) between reference and measured readings. The 95% CI for the mean difference between predicted and measured diastolic BP was ±0.57 (-0.41 to 0.73) mm Hg. Systolic BP predictions yielded a mean absolute percentage error of 9.52% and a mean difference of 2.69 mm Hg (SD 7.86 mm Hg). The 95% CI for the mean difference between predicted and measured systolic BP was ±1.14 (-1.54 to -3.83) mm Hg. Hemoglobin concentration predictions had a mean absolute percentage error of 8.52%, with a mean difference of 0.23 g/dL (SD 0.67 g/dL). The 95% CI for the mean difference between predicted and reference measured hemoglobin concentration was ±0.10 (95% CI 0.13-0.33) g/dL.

CONCLUSIONS

Noninvasive rPPG-based measurement of BP and hemoglobin concentration at the preoperative evaluation setting has great potential for improving convenience, improving efficiency, and conserving resources for patients and health care providers. Our model was able to accurately predict diastolic BP in patients with diverse skin tones and medical comorbidities. The findings of this study serve as a basis for further studies to develop and validate the model for noninvasive rPPG-based BP and hemoglobin concentration measurement.

Для цитирования: Tan Selene Y L, Chai Jia Xin, Choi Minwoo, Javaid Umair, Tan Brenda Pei Yi, Chow Belinda Si Ying, Abdullah Hairil Rizal. Remote Photoplethysmography Technology for Blood Pressure and Hemoglobin Level Assessment in the Preoperative Assessment Setting: Algorithm Development Study.. JMIR formative research. 2025. DOI: 10.2196/60455.