MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1159/000519713

Медицинская библиотека · Neonatology · 2021

Comparison of Cardiac Output Measurement by Electrical Velocimetry with Echocardiography in Extremely Low Birth Weight Neonates.

Hassan Mohammad Ahmad, Bryant Manuel B, Hummler Helmut D

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Electrical velocimetry (EV) offers a noninvasive tool for continuous cardiac output (CO) measurements which might facilitate hemodynamic monitoring and targeted therapy in low birth neonates, in whom other methods of CO measurement are not practicably feasible.

METHODS

This prospective observational study compared simultaneous cardiac output measurements by electrical velocimetry (COEV) with transthoracic echocardiography (COTTE) in extremely low birth weight (ELBW) neonates in the neonatal intensive care unit (NICU). Echocardiography was performed by 1 single examiner. Data were analyzed by Bland-Altman analysis and independent-samples analysis of variance. A mean percentage error (MPE) of <30% and limits of agreement (LOA) up to ±30% were considered clinically acceptable.

RESULTS

Thirty-eight ELBW neonates were studied and yielded 85 pairs of COEV and COTTE measurements. Bland-Altman analysis showed an overall bias (the mean difference) and LOA of -126 and -305 to +52 mL min-1, respectively, and an MPE of 66%. Patients with patent ductus arteriosus had a higher bias with LOA and MPE of -166.8, -370.7 to +37 mL min-1, and 69%, respectively. The overall true precision was 58%.

CONCLUSION

This study showed high bias and lack of agreement between EV and TTE for measurement of CO in ELBW infants in NICU, limiting applicability of EV to monitor absolute values.

Для цитирования: Hassan Mohammad Ahmad, Bryant Manuel B, Hummler Helmut D. Comparison of Cardiac Output Measurement by Electrical Velocimetry with Echocardiography in Extremely Low Birth Weight Neonates.. Neonatology. 2021. DOI: 10.1159/000519713.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.