MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tbme.2025.3623609

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2026

FOSTER: A Comprehensive Pipeline for Transabdominal Fetal Pulse Oximetry Validated in a Large Animal Model of Pregnancy.

Kasap Begum, Saffarpour Mahya, Qian Weitai, Joarder Rishad Raiyan, Vali Kourosh, Ghiasi Soheil

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Transabdominal fetal pulse oximetry (TFO) has the potential to supplement present intrapartum fetal monitoring approaches, which cannot accurately detect fetuses at risk of birth asphyxia. However, non-invasive measurement of fetal oxygen saturation (fSpO$_{2}$) is challenging due to dominant maternal tissue signals. We present methods to overcome such challenges, enabling robust and continuous fSpO$_{2}$ measurement.

METHODS

We introduce FOSTER (Fetal Oxygen SaTuration EstimatoR), a comprehensive pipeline that combines novel signal processing and machine learning techniques to process photoplethysmogram (PPG) signals and estimate continuous fSpO$_{2}$. Using controlled desaturation experiments in pregnant ewes, we evaluate FOSTER's performance with both mixed maternal-fetal signals and isolated fetal components.

RESULTS

Processing isolated fetal signals improves estimation accuracy with respect to arterial blood oxygen saturation (SaO$_{2}$), showing improvements of 13.7% in mean absolute error (MAE) and 10.5% in Pearson correlation coefficient relative to using mixed TFO signals alone. A dual-branch neural network, processing mixed and isolated PPG signals simultaneously, achieves additional improvements of 6.3% in MAE and 4.1% in Pearson correlation compared to using isolated fetal signals alone.

CONCLUSION

The FOSTER pipeline demonstrates significant improvements in continuous fSpO$_{2}$ estimation accuracy through advanced signal processing and a dual-branch architecture, establishing a foundation for reliable fSpO$_{2}$ monitoring.

SIGNIFICANCE: This work represents an important innovative step toward accurate, continuous, and non-invasive monitoring of fetal oxygenation. The validated methods in animal studies establish a foundation for advancing the development of fetal monitoring systems, offering new possibilities for improved maternal-fetal care.

Для цитирования: Kasap Begum, Saffarpour Mahya, Qian Weitai, Joarder Rishad Raiyan, Vali Kourosh, Ghiasi Soheil. FOSTER: A Comprehensive Pipeline for Transabdominal Fetal Pulse Oximetry Validated in a Large Animal Model of Pregnancy.. IEEE transactions on bio-medical engineering. 2026. DOI: 10.1109/tbme.2025.3623609.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.