Aerosol Delivery With a Vibrating Mesh Nebulizer Across Tidal Volume-Based Pediatric Invasive Ventilation Models: An In Vitro Evaluation.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Miura Shinya, Igarashi Yoshihiro, Shono Kenta, Shima Saho, Kawaguchi Atsushi

Аннотация

CONTEXT: Despite the widespread use of vibrating mesh nebulizers (VMNs) for aerosol therapy during mechanical ventilation, there is limited pediatric evidence on how tidal volume (TV)-based ventilation models and commonly modified ventilator- and circuit-related factors affect aerosol delivery.

HYPOTHESIS: We hypothesized that these ventilator- and circuit-related modifications would affect the efficacy of aerosol delivery with a VMN.

METHODS AND MODELS: We conducted an in vitro experiment using six pediatric ventilation models with TVs set at 8 mL/kg (TV: 40-400 mL for 5-50 kg), representing children from infancy to adolescence. We nebulized salbutamol with a VMN positioned on the dry side of a heated humidifier, and quantified inhaled mass fraction by measuring the drug collected on a filter positioned distal to the endotracheal tube using a spectrophotometer. We evaluated the effects of TV-based models and variations in other settings, including endotracheal tube sizes, bias flows, inspiratory rise time, ventilator modes, circuit configurations, VMN positions, and drug concentrations.

RESULTS: In the baseline model with a TV of 80 mL, the mean ± sd of inhaled mass fraction was 11.2% ± 0.3%. It increased progressively across the TV-based models: 3.5% ± 0.2% (TV: 40 mL), 7.9% ± 0.4% (TV: 56 mL), 15.7% ± 0.2% (TV: 120 mL), 19.4% ± 0.4% (TV: 160 mL), and 21.3% ± 1.7% (TV: 400 mL; all p < 0.001 vs. baseline). No significant differences were observed in variations of the other ventilatory, endotracheal tube-, or circuit-related settings.

INTERPRETATION AND CONCLUSIONS: VMN aerosol delivery increased across TV-based ventilation models and was minimally affected by other ventilator settings, endotracheal tube size, and circuit-related modifications. Consistent aerosol delivery at a given TV suggests that TV is the dominant determinant of aerosol delivery efficacy.