DOI 10.1136/archdischild-2023-326278

Медицинская библиотека · Archives of disease in childhood. Fetal and neonatal edition · 2024

Comparison of two different oxygen saturation target ranges for automated oxygen control in preterm infants: a randomised cross-over trial.

Brouwer Fleur, Salverda Hylke H, Cramer Sophie J E, Schmeits Chantal, van der Plas Jacoline, Te Pas Arjan B, Dekker Janneke

СтатьяРандомизированное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To compare the effect of peripheral oxygen saturation (SpO2) target range (TR) (either 91%-95% and 92%-96%) on the frequency and duration of hypoxic and hyperoxic episodes while on automated oxygen control using the OxyGenie controller.

DESIGN

Randomised cross-over study.

SETTING

Tertiary-level neonatal unit in the Netherlands.

PATIENTS: Infants (n=27) with a median (IQR) gestational age of 27+0 (25+5-27+3) weeks and postnatal age of 16 (10-22) days, receiving invasive or non-invasive respiratory support.

INTERVENTIONS: In both groups supplemental oxygen was titrated to a TR of 91%-95% (TRlow) or 92%-96% (TRhigh) by the OxyGenie controller (SLE6000 ventilator) for 24 hours each, in random sequence. After a switch in TR, a 1-hour washout period was applied to prevent carry-over bias.

MAIN OUTCOME MEASURES: Frequency and duration of hypoxic (SpO2<80% for ≥1 s) and hyperoxic episodes (SpO2>98% for ≥1 s).

RESULTS

Hypoxic episodes were less frequent when the higher range was targeted (TRhigh vs TRlow: 2.5 (0.7-6.2)/hour vs 2.4 (0.9-10.2)/hour, p=0.02), but hyperoxic episodes were more frequent (5.3 (1.8-12.3)/hour vs 2.9 (1.0-7.1)/hour, p<0.001). The duration of the out-of-range episodes was not significantly different (hypoxia: 4.7 (2.8-7.1) s vs 4.4 (3.7-6.5) s, p=0.67; hyperoxia: 4.3 (3.3-4.9) s vs 3.9 (2.8-5.5) s, p=0.89).

CONCLUSION

Targeting a higher SpO2 TR with the OxyGenie controller reduced hypoxic episodes but increased hyperoxic episodes. This study highlights the feasibility of using an automated oxygen titration device to explore the effects of subtle TR adjustments on clinical outcomes in neonatal care.

TRIAL REGISTRATION NUMBER: NL9662.

Для цитирования: Brouwer Fleur, Salverda Hylke H, Cramer Sophie J E, Schmeits Chantal, van der Plas Jacoline, Te Pas Arjan B, Dekker Janneke. Comparison of two different oxygen saturation target ranges for automated oxygen control in preterm infants: a randomised cross-over trial.. Archives of disease in childhood. Fetal and neonatal edition. 2024. DOI: 10.1136/archdischild-2023-326278.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.