DOI 10.1097/scs.0000000000001736

Медицинская библиотека · The Journal of craniofacial surgery · 2015

Mechanical Analysis of Ultrasound-Activated Pins and Resorbable Screws: Two Different Techniques to Fixate Osteosynthesis in Craniosynostosis Surgery.

Savolainen Mikko J, Ritvanen Antti G, Koljonen Virve S, Turunen Markus P K, Hulkkonen Hanna H, Vuorinen Vesa H O, Leikola Junnu P

СтатьяПеревод в очереди

Аннотация

PURPOSE: Ultrasound activation of resorbable pins directly into drilled holes of the calvarium was introduced to overcome the time-consuming installation in the resorbable osteosynthesis fixation in craniosynostosis surgery. There is paucity in the data comparing the mechanical properties of resorbable screws and ultrasound-activated pins produced by different manufacturers. The aim of this experimental study was to compare the mechanical properties of ultrasound-activated pins and resorbable screws.

METHODS

A mechanical testing machine was used to characterize the mechanical performance of screws and ultrasound pins. The screws and pins were tested individually in 2 directions with respect to the longitudinal axis: vertical, that is, axial pull-out strength and horizontal, that is, shear strength. The mean maximum strength of fixation was determined. Broken screws and pinheads were analyzed by a scanning electron microscope to determine the site of fracture.

RESULTS

All of the resorbable screws and pins broke at the point where the device enters bone. In pull-out testing, the mean maximum strength of the ultrasound-activated pins was 30.5 ± 5.4 N and that of the resorbable screws was 54.0 ± 0.3 N. In shear testing, the mean maximum strength of ultrasound-activated pins was 57.1 ± 20.1 N and that of the resorbable screws was 53.9 ± 0.4 N.

CONCLUSIONS

In their intended configuration, there is no clinically significant difference in fixation strength between ultrasound-activated pins and resorbable screws.

Для цитирования: Savolainen Mikko J, Ritvanen Antti G, Koljonen Virve S, Turunen Markus P K, Hulkkonen Hanna H, Vuorinen Vesa H O, Leikola Junnu P. Mechanical Analysis of Ultrasound-Activated Pins and Resorbable Screws: Two Different Techniques to Fixate Osteosynthesis in Craniosynostosis Surgery.. The Journal of craniofacial surgery. 2015. DOI: 10.1097/scs.0000000000001736.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.