MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2014.04.013

Медицинская библиотека · JACC. Cardiovascular interventions · 2014

Radiation dose benchmarks during cardiac catheterization for congenital heart disease in the United States.

Ghelani Sunil J, Glatz Andrew C, David Sthuthi, Leahy Ryan, Hirsch Russel, Armsby Laurie B, Trucco Sara M, Holzer Ralf J, Bergersen Lisa

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The aim of this study was to define age-stratified, procedure-specific benchmark radiation dose levels during interventional catheterization for congenital heart disease.

BACKGROUND

There is a paucity of published literature with regard to radiation dose levels during catheterization for congenital heart disease. Obtaining benchmark radiation data is essential for assessing the impact of quality improvement initiatives for radiation safety.

METHODS

Data were obtained retrospectively from 7 laboratories participating in the Congenital Cardiac Catheterization Project on Outcomes collaborative. Total air kerma, dose area product, and total fluoroscopy time were obtained for the following procedures: 1) patent ductus arteriosus closure; 2) atrial septal defect closure; 3) pulmonary valvuloplasty; 4) aortic valvuloplasty; 5) treatment of coarctation of aorta; and 6) transcatheter pulmonary valve placement.

RESULTS

Between January 2009 and July 2013, 2,713 cases were identified. Radiation dose benchmarks are presented including median, 75th percentile, and 95th percentile. Radiation doses varied widely between age groups and procedure types. Radiation exposure was lowest in patent ductus arteriosus closure and highest in transcatheter pulmonary valve placement. Total fluoroscopy time was a poor marker of radiation exposure and did not correlate well with total air kerma and dose area product.

CONCLUSIONS

This study presents age-stratified radiation dose values for 6 common congenital heart interventional catheterization procedures. Fluoroscopy time alone is not an adequate measure for monitoring radiation exposure. These values will be used as baseline for measuring the effectiveness of future quality improvement activities by the Congenital Cardiac Catheterization Project on Outcomes collaborative.

Для цитирования: Ghelani Sunil J, Glatz Andrew C, David Sthuthi, Leahy Ryan, Hirsch Russel, Armsby Laurie B, Trucco Sara M, Holzer Ralf J, Bergersen Lisa. Radiation dose benchmarks during cardiac catheterization for congenital heart disease in the United States.. JACC. Cardiovascular interventions. 2014. DOI: 10.1016/j.jcin.2014.04.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.