The optimal method for recording prosthetic heart valve sounds in clinical situations.

Bito Yasuyuki, Soeta Yoshiharu, Nakagawa Seiji, Shibata Toshihiko, Suehiro Shigefumi

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Valve sounds are thought to be useful in evaluating the functioning of prosthetic heart valves. In previous reports, two recording instruments have mainly been used, a condenser microphone (instrument A) and an accelerometer (instrument B), respectively. This study aimed to investigate the applicability of these conventional devices in recording mechanical heart valve sounds in daily clinical situations. In addition, two other instruments, namely, a stethoscope including a small microphone in its tube (instrument C) and an electrostethoscope (instrument D), were proposed and examined.

METHODS

Mechanical valve sounds were recorded from postoperative patients using each-device and the signal to noise ratio (SNR) of each was calculated. Frequency characteristics of the two newly proposed instruments were also investigated by recording a sweep signal in an anechoic chamber.

RESULTS

The SNR values of the new instruments were significantly superior to those of conventional ones (A, sound signals were not defined from noises; B, 0.97 +/- 0.59 dB; C, 11.5 +/- 4.7 dB; and D, 9.4 +/- 3.5 dB, respectively p<0.05). The frequency characteristics of the two newly proposed instruments were also shown to be acceptable (under 10 kHz), though resonance frequency peaks were observed.

CONCLUSIONS

Two newly proposed instruments can record mechanical heart valve sounds more noiselessly and with more ease than previous ones, and their frequency characteristics are acceptable.

Для цитирования: Bito Yasuyuki, Soeta Yoshiharu, Nakagawa Seiji, Shibata Toshihiko, Suehiro Shigefumi. The optimal method for recording prosthetic heart valve sounds in clinical situations.. Osaka city medical journal. 2005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.