MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1553350615617250

Медицинская библиотека · Surgical innovation · 2015

Feasibility of an Intraluminal Bypass Device in Low Colorectal Anastomosis: Preliminary Results in a Porcine Model.

Willenz Udi, Wasserberg Nir, Botero-Anug Ana Maria, Greenberg Ron

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The Cologuard CG-100 is a novel intraluminal bypass device designed to reduce the clinical outcomes associated with low colorectal anastomotic leak. The device is inserted transanally, anchored to the colon above the anastomosis, and deployed intraluminally to cover the anastomosis from within. The purpose of this study was to evaluate the safety and performance of the device in a porcine model.

METHOD: Twelve pigs underwent low colorectal anastomosis with insertion of the Cologuard CG-100 device. Contrast material injection, abdominal X-ray, and histologic studies were used to evaluate sealing quality, device positioning, and tissue damage, respectively. The surgeons completed a usability and satisfaction questionnaire after completion of the procedure.

RESULTS

Absolute sealing was observed in all 4 animals euthanized immediately after surgery. In the other 8 animals, the device was kept in situ for 10 days and then extracted. X-ray films with injection of contrast material through a designated injection tube before device removal showed that the sheath and ring were correctly placed. No leak was demonstrated. There were no device-related adverse events, and no critical histological abnormalities were noted in the bowel area that was compressed by the device. The device was found to be easy to insert, position, and extract.

CONCLUSION

The Cologuard CG-100 device efficiently reduced contact between fecal content and low colorectal anastomosis in a porcine model and is easily deployed and extracted. It holds promise for possible clinical use pending further studies.

Для цитирования: Willenz Udi, Wasserberg Nir, Botero-Anug Ana Maria, Greenberg Ron. Feasibility of an Intraluminal Bypass Device in Low Colorectal Anastomosis: Preliminary Results in a Porcine Model.. Surgical innovation. 2015. DOI: 10.1177/1553350615617250.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.