DOI 10.1097/pcc.0000000000003282

Медицинская библиотека · Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies · 2023

Pre-Extubation Cuffed Tube Leak Test and Subsequent Post-Extubation Laryngeal Edema: Prospective, Single-Center Evaluation of PICU Patients.

Kanno Koji, Fujiwara Naoki, Moromizato Takuhiro, Fujii Shuichi, Ami Yuki, Tokushige Akihiro, Ueda Shinichiro

Наблюдательное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

We performed our standard air leak, leak percentage, and cuff leak percentage tests in pediatric patients intubated with microcuff pediatric tracheal tubes (MPTTs) just before extubation. We examined the association between test findings and the subsequent occurrence of post-extubation laryngeal edema (PLE).

DESIGN

Prospective, single-center, observational study.

SETTING

PICU (June 1, 2020 to May 31, 2021).

PATIENTS: Pediatric patients intubated and scheduled for extubation during the day shift in the PICU.

INTERVENTIONS: Multiple pre-extubation leak tests were performed on each patient immediately before extubation. In our center, the standard leak test is positive if a leak is audible at 30 cm H 2 O applied pressure with the MPTT cuff deflated. Two other tests were calculated in the pressure control-assist control ventilator mode using the following formulas: leak percentage with deflated cuff = (inspiratory tidal volume [V t ]-expiratory V t ) × 100/inspiratory V t ; cuff leak percentage = (expiratory V t with inflated cuff-expiratory V t with deflated cuff) × 100/expiratory V t with inflated cuff.

MEASUREMENTS AND MAIN

RESULTS

The diagnostic criteria for PLE was made by at least two healthcare professionals and included upper airway stricture with stridor-requiring nebulized epinephrine. Eighty-five pediatric patients (< 15 yr) who had been intubated for at least 12 hours using the MPTT were included. Positive rates for the standard leak, leak percentage (cutoff 10%), and cuff leak percentage (cutoff 10%) tests were 0.27, 0.20, and 0.64, respectively. The standard leak, leak percentage, and cuff leak tests showed sensitivities of 0.36, 0.27, and 0.55, respectively; and specificities of 0.74, 0.81, and 0.35, respectively. PLE occurred in 11 of 85 patients (13%), and there were no instances of needing reintubation.

CONCLUSIONS

The pre-extubation leak tests in current practice for intubated pediatric patients in the PICU all lack diagnostic accuracy for PLE.

Для цитирования: Kanno Koji, Fujiwara Naoki, Moromizato Takuhiro, Fujii Shuichi, Ami Yuki, Tokushige Akihiro, Ueda Shinichiro. Pre-Extubation Cuffed Tube Leak Test and Subsequent Post-Extubation Laryngeal Edema: Prospective, Single-Center Evaluation of PICU Patients.. Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies. 2023. DOI: 10.1097/pcc.0000000000003282.