Accuracy of analog telephonic stethoscopy for pediatric telecardiology.

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка службы здравоохранения СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Belmont John M, Mattioli Leone F

Аннотация

OBJECTIVE: Wide-bandwidth electronic stethoscopy is reliable and accurate for pediatric telecardiology. We tested a much less expensive and more convenient system for the same purpose, a narrow-bandwidth telephonic stethoscope (TS).

METHODS: Seventy-six consecutive patients (mean age: 10.0; standard deviation: 6.5 years) in a pediatric cardiology outpatient clinic were studied. One pediatric cardiologist examined the patients with his acoustic stethoscope (AS); a second examined them within a few minutes using a remote TS. A nurse placed the TS chest piece as directed by the remote examiner via intercom, but neither video examination nor conversation with the parent/patient were permitted. Examiners independently recorded the stethoscope findings for all heart sounds, all murmurs, and heart disease (present/absent). TS accuracy was indexed using the kappa statistic for TS/AS agreement and for TS agreement with auscultatory findings predicted from echocardiographic (echo) studies (N = 49).

RESULTS: TS/AS agreement was satisfactory for presence/absence of heart disease (kappa = 0.63) and for organic, functional, vibratory, diastolic aortic, and diastolic pulmonic murmurs (kappa range: 0.65-0.75). For other specific murmurs and all heart sounds, TS/AS agreement was either unsatisfactory (kappa < or = 0.60) or indeterminate because prevalence was 0. TS-AS agreement improved when the TS was used by the more-experienced TS examiner and with patients at least 5 years of age. When the older children were examined by the more TS-experienced examiner, the TS-echo comparison yielded kappa = 0.90, raw agreement = 0.96, sensitivity = 0.94, and specificity = 1.00.

CONCLUSIONS: In pediatric patients, a narrow-bandwidth telephonic stethoscope can accurately distinguish between functional and organic murmurs and thus can detect heart disease. Accuracy is greatest when the instrument is used by an experienced examiner with patients at least 5 years of age.