DOI 10.1002/ppul.24589

Медицинская библиотека · Pediatric pulmonology · 2019

In vitro comparison of unit dose vs infusion pump administration of albuterol via high-flow nasal cannula in toddlers.

Li Jie, Wei Wu, Fink James B

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

Transnasal pulmonary aerosol delivery using high-flow nasal cannula (HFNC) devices has become a popular route of aerosol administration in toddlers. Clinically, albuterol is administered using an infusion pump or unit doses. However, little evidence is available to compare the two administration strategies.

METHODS

A toddler manikin (15 kg) with appropriate anatomic airway was connected with collecting filter to a simulator of distressed breathing. HFNC device with mesh nebulizer placed at the inlet of a humidifier at 37°C, with the gas flow set at 25 and 3.75 L/min. Five milligrams of albuterol was delivered in all experiments. With infusion pump administration, albuterol concentrations of 5 and 1 mg/mL were delivered at 4 and 20 mL/hr for 15 minutes. With unit dose administration, 1 mL (5 mg/mL) and 2 mL (2.5 mg/mL) of albuterol were nebulized. Additional tests with mouth open and nebulizers via mask were using 5 mg/1 mL for mesh nebulizer and 5 mg/3 mL for jet nebulizer (n = 3). The drug was eluted from the filter and assayed with UV spectrophotometry (276 nm).

RESULTS

The inhaled dose was higher with unit dose than infusion pump administration with gas flows of 25 L/min (2.66 ± 0.38 vs 1.16 ± 0.28%; P = .004) and 3.75 L/min (10.51 ± 1.29 vs 8.58 ± 0.68%; P = .025). During unit dose administration, compared with closed-mouth breathing, open-mouth breathing generated a higher inhaled dose at 3.75 L/min and lower inhaled dose at 25 L/min. Compared to the nebulizers via mask with both open and closed-mouth breathing, nebulization via HFNC at 3.75 L/min generated greater inhaled dose, while HFNC at 25 L/min generated lower inhaled dose.

CONCLUSIONS

During transnasal aerosol delivery, the inhaled dose was higher with medication administrated using unit dose than using an infusion pump.

Для цитирования: Li Jie, Wei Wu, Fink James B. In vitro comparison of unit dose vs infusion pump administration of albuterol via high-flow nasal cannula in toddlers.. Pediatric pulmonology. 2019. DOI: 10.1002/ppul.24589.