DOI 10.1016/j.ejmp.2024.104840
Predicting radiotoxic effects after BNCT for brain cancer using a novel dose calculation model.
Аннотация
PURPOSE: The normal brain is an important dose-limiting organ for brain cancer patients undergoing radiotherapy. This study aims to develop a model to calculate photon isoeffective doses (DIsoE) to normal brain that can explain the incidence of grade 2 or higher somnolence syndrome (SS⩾2) after Boron Neutron Capture Therapy (BNCT).
METHODS
A DIsoE model was constructed to find the reference photon dose that equals the Normal Tissue Complication Probability (NTCP) of the absorbed dose from BNCT. Limb paralysis rates from the rat spinal cord model exposed to conventional or BNCT irradiation were used to determine model parameters. NTCP expressions for both irradiations were constructed based on Lyman's model accordingly. DIsoE values were calculated for BNCT treatments performed in Finland and USA. An equivalent uniform dose (EUD) based on peak and average whole-brain doses and treatment fields was also introduced. Combining DIsoE and EUD models, a dose-response curve for SS⩾2 in BNCT patients was constructed and compared to conventional radiotherapy outcomes.
RESULTS
The DIsoE model reveals higher than expected photon-equivalent doses in the brain, indicating the need to modify standard dose calculation methods. Neither peak dose nor average whole-brain dose alone predicts SS⩾2 development. However, the dose-response curve derived from combining DIsoE and EUD models effectively explains the incidence of SS⩾2 after BNCT.
CONCLUSIONS
The introduced DIsoE and EUD models predict the incidence of somnolence syndrome after BNCT. The first dose-response relationship for SS⩾2 derived entirely from brain tumour patients treated with BNCT, consistent with photon radiotherapy responses, is presented.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.