MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.surg.2004.06.040

Медицинская библиотека · Surgery · 2004

Intraoperative electromyographic monitoring of the recurrent laryngeal nerve in reoperative thyroid and parathyroid surgery.

Yarbrough Donald E, Thompson Geoffrey B, Kasperbauer Jan L, Harper C Michel, Grant Clive S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Injury to the recurrent laryngeal nerve (RLN) is a rare complication of initial thyroid and parathyroid surgery, but the prevalence is much higher in the reoperative setting. The use of continuous, intraoperative electromyographic monitoring of the RLN has been suggested to improve the safety of cervical explorations.

METHODS

Outcomes of a group of reoperative thyroid and parathyroid cases that used EMG monitoring with endoscopically applied hook-wire electrodes were compared with a group of cervical reoperations without monitoring. Office laryngoscopy (indirect or fiberoptic) was used to evaluate and follow suspected RLN complications.

RESULTS

Electromyography was used in 52 cervical reexploration procedures. Patients averaged 1.8 previous explorations (range, 1-7 explorations) and underwent procedures for parathyroid (31%) and/or thyroid (77%) disease (overall, 72% malignant). The non-monitored group had 59 patients with similar characteristics. Only 1 permanent nerve complication in each group was unintended (electromyography, 1.9%; non-electromyography, 1.7%). Seven false-negative and 2 false-positive electromyographic findings occurred. No complications resulted from placement of the electromyography electrodes.

CONCLUSIONS

Intraoperative electromyographic monitoring of the RLN in reoperative neck surgery can be performed safely but did not decrease RLN complications in this study. Experience and routine nerve exposure remain crucial to the minimization of RLN complications.

Для цитирования: Yarbrough Donald E, Thompson Geoffrey B, Kasperbauer Jan L, Harper C Michel, Grant Clive S. Intraoperative electromyographic monitoring of the recurrent laryngeal nerve in reoperative thyroid and parathyroid surgery.. Surgery. 2004. DOI: 10.1016/j.surg.2004.06.040.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.