MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3171/2025.2.peds24622

Медицинская библиотека · Journal of neurosurgery. Pediatrics · 2025

Complex shunt system comparison: an observational study by the Hydrocephalus Clinical Research Network.

Jackson Eric M, Jensen Hailey, Reeder Ron W, Chiarelli Peter A, Chu Jason, Hankinson Todd C, Hauptman Jason S, Isaacs Albert M, Kulkarni Abhaya V, Limbrick David D, McDonald Patrick J, Pindrik Jonathan A, Pollack Ian F, Rocque Brandon G, Rozzelle Curtis J, Strahle Jennifer M, Tamber Mandeep S, Wellons John C, Kestle John R W, Whitehead William E, Riva-Cambrin Jay

СтатьяНаблюдательное исследованиеСравнительное исследованиеМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Treatment of multiloculated hydrocephalus can require multiple procedures and shunt catheters. This study aimed to determine whether there are differences in shunt survival or complications in patients with complex shunt systems based on whether they have separate systems or multiple intracranial catheters with a single distal catheter and a Y- or T-connector.

METHODS

The authors retrospectively reviewed the Hydrocephalus Clinical Research Network (HCRN) Core Data Project, a prospective multicenter registry. Patients undergoing first-time placement of a complex ventriculoperitoneal shunt or first-time conversion to a complex shunt were identified and included in the analysis. Propensity-weighted Cox regression was used to control for HCRN center and etiology using the rate of shunt failure (shunt malfunction or shunt infection) as the primary outcome. The final regression model was also adjusted for age and complex chronic conditions.

RESULTS

In total, 369 patients were included. One hundred fifty-one patients had separate systems and 218 had Y/T-connectors. After adjustment for age and comorbidities, the rate of shunt failure for systems with Y/T-connectors was not significantly different than that for separate shunt systems: 62% versus 55% (HR 1.20, 95% CI 0.91-1.59, p = 0.197). There was a statistically significant difference in operative time with separate systems having shorter operative times (mean time 63.0 vs 80.0 minutes; mean difference 16.32, 95% CI 7.53-25.10, p < 0.001).

CONCLUSIONS

There were no differences in the shunt failure rates or complications between the separate shunt systems and Y/T-connector systems used to treat complex shunts. However, surgical time was significantly shorter with separate shunt systems. These findings suggest that surgeons can tailor the shunt system on the basis of individual patient characteristics.

Для цитирования: Jackson Eric M, Jensen Hailey, Reeder Ron W, Chiarelli Peter A, Chu Jason, Hankinson Todd C, Hauptman Jason S, Isaacs Albert M, Kulkarni Abhaya V, Limbrick David D, McDonald Patrick J, Pindrik Jonathan A, Pollack Ian F, Rocque Brandon G, Rozzelle Curtis J, Strahle Jennifer M, Tamber Mandeep S, Wellons John C, Kestle John R W, Whitehead William E, Riva-Cambrin Jay. Complex shunt system comparison: an observational study by the Hydrocephalus Clinical Research Network.. Journal of neurosurgery. Pediatrics. 2025. DOI: 10.3171/2025.2.peds24622.

Предметные рубрики MeSH

Подготовка и перевод публикации