DOI 10.1093/chromsci/47.8.689

Медицинская библиотека · Journal of chromatographic science · 2009

Optimization of a novel procedure for determination of VOCs in water and human urine samples based on SBSE coupled with TD-GC-HRMS.

Jakubowska Natalia, Henkelmann Bernhard, Schramm Karl-Werner, Namiesnik Jacek

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

In this study, stir-bar sorptive extraction and thermal desorption followed by gas chromatography coupled with high resolution mass spectrometry was applied for determination of halo-organic compounds (bromodichloromethane, dibromochloromethane, bromoform, and tetrachloroethylene) in water and human urine samples. Time of extraction and stirring speed were optimized. The results show that the optimum extraction time is 30 min with 600 rpm of stirring speed with Twister of 20 mm in length and 1.0-mm film thickness of PDMS (126 microL). The calibration curves, limits of detection and quantification for all compounds were calculated. This procedure is characterized by very low limits of detection and quantitation: lower than 0.0017 microg/L and good repeatability for all four volatile compounds. This new analytical procedure was identified to be easy, reliable, sensitive, and requires only small amounts of sample. It can constitute a good alternative to well-known procedures based on application of head space and gas chromatography coupled with electron capture detection.

Для цитирования: Jakubowska Natalia, Henkelmann Bernhard, Schramm Karl-Werner, Namiesnik Jacek. Optimization of a novel procedure for determination of VOCs in water and human urine samples based on SBSE coupled with TD-GC-HRMS.. Journal of chromatographic science. 2009. DOI: 10.1093/chromsci/47.8.689.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.