Comparison of Arterial Partial Pressure of Carbon Dioxide with That at Oxygenator Exhaust During Cardiopulmonary Bypass for Pediatric Cardiac Surgery.

СтатьяНаблюдательное исследованиеСравнительное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Devarakonda Bhargava V, Gadhinglajkar Shrinivas V, Vadapalli Kiranmai, Sukesan Subin, Sreedhar Rupa, Dharan Baiju S

Аннотация

OBJECTIVE: To evaluate whether carbon dioxide tension (PeCO₂) derived from oxygenator exhaust capnography is comparable to temperature-corrected arterial partial pressure of carbon dioxide (PaCO₂) measured via blood gas analysis during mild hypothermic cardiopulmonary bypass (CPB) in pediatric patients undergoing elective cardiac surgery.

DESIGN: A prospective observational study.

SETTING: A tertiary care hospital with a specialized pediatric cardiac surgery unit.

PARTICIPANTS: Eighty paired measurements of PeCO₂ and PaCO₂ were obtained from 40 pediatric patients (aged between 1 and 12 years) undergoing elective cardiac surgery for biventricular repair with stable mild hypothermic CPB using a membrane oxygenator.

INTERVENTIONS: No interventions were performed. The study involved continuous measurement of PeCO₂ from the oxygenator exhaust and periodic arterial blood gas analysis to assess PaCO₂ during stable mild hypothermia.

MEASUREMENTS AND MAIN RESULTS: A Bland-Altman analysis was conducted to determine the agreement between PeCO₂ and temperature-corrected PaCO₂. The correlation coefficient was 0.81, with a bias of 0.036, standard deviation of 2.76, and limits of agreement ranging from -5.37 to 5.44. A linear regression equation was established: temperature-corrected PaCO₂ = 0.71 × PeCO₂ + 8.28. No significant correlation was observed between PeCO₂ and temperature-uncorrected PaCO₂.

CONCLUSIONS: Under stable mild hypothermic CPB conditions, PeCO₂ demonstrates a strong correlation with temperature-corrected PaCO₂ and can be predicted using a simple linear equation in pediatric patients undergoing elective cardiac surgery. Oxygenator exhaust capnography may serve as a continuous, cost-effective, non-invasive, pragmatic surrogate for PaCO₂ monitoring during a stable hypothermic CPB period in the pediatric population undergoing elective cardiac surgery.