Diplopia in orbital fractures: a simple method to evaluate eye motility.
Аннотация
CONCLUSION
This study on eye motility demonstrates a new application of electro-oculography which proved to have a high potential for establishing and diagnosing the cause of diplopia in orbital floor fractures. Electro-oculography is objective and easy to perform even when the restricted eye is covered. The 'uninjured' eye can be used as the reference. Any motility difference between the left and the right eye was easily detected.
OBJECTIVE
It is essential in orbital floor fractures to distinguish entrapment from other causes of diplopia. We investigated whether vertical eye motility and restricted elevation can be detected and characterized by means of electro-oculography.
MATERIALS AND
METHODS
Electronystagmography equipment used for vestibular testing in daily ENT practice was used in a different context. Unilateral mechanical restriction was generated in the test subjects through fixation of one eye bulb. The velocity and range of eye motility of both eyes were measured and compared in 12 healthy test subjects and 7 patients with long-term vertical diplopia after orbital floor fractures.
RESULTS
Simulated entrapment was easily detected as a distinct difference between right and left eye motility recordings (p<0.001). Recordings of patients with diplopia differed significantly from those of healthy test subjects with respect to velocity (p<0.05) and range of motility (p<0.05).
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.