Percutaneous closure of perimembranous ventricular septal defects utilizing almost ideal Amplatzer Duct Occluder II: Why limitation in sizes?
Аннотация
AIM: The purpose of this study is to describe the special aspects of perimembranous ventricular septal defects (pmVSD) closure by utilizing Amplatzer Duct Occluder II (ADO II) devices with a rational request for bigger ADO-II sizes, based on our experience in transcatheter device closure of pmVSD.
METHODS AND
RESULTS
At our institution, placement of an ADO II device was used in 15 patients with pmVSD; the patients' age ranged between 6 months and 20 years. The indications for closure were CHF (n = 4), hemodynamically significant shunt (n = 7), tricuspid regurgitation (n = 3), and high risk for infective endocarditis (n = 2), respectively. The location of the VSD was infracristal in 13 patients, supracristal in 1, and a postsurgical Gerbode VSD in another one. Implantation of the device was successfully performed without embolization, any evidence of an AV block, or other conductance abnormalities during implantation and follow-up in the mean of 2.5 years (range 2 months-6.5 years).
CONCLUSIONS
Transcatheter closure of a pmVSD with ADO II is feasible in all pediatric and young adult age groups, by considering the device diameter limitations. The off-label use of ADO II implantation seems to be safe for VSDs closure up to 6 mm of size and feasible for various locations including unusual morphology such as postsurgical Gerbode defect.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.