DOI 10.1186/s13019-018-0823-5

Медицинская библиотека · Journal of cardiothoracic surgery · 2019

Vibration transmittance measures sternotomy stability - a preliminary study in human cadavers.

Hautalahti Juha, Joutsen Atte, Goebeler Sirkka, Luukkaala Tiina, Khan Jahangir, Hyttinen Jari, Laurikka Jari

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Stability is essential for the normal healing of a sternotomy. Mechanical vibration transmittance may provide a new means of early detection of diastasis in the sternotomy and thus enable the prevention of further complications. We sought to confirm that vibration transmittance detects sternal diastasis in human tissue.

METHODS

Ten adult human cadavers (8 males and 2 females) were used for sternal assessments with a device constructed in-house to measure the transmittance of a vibration stimulus across the median sternotomy at the second, third, and fourth costal cartilage. Intact bone was compared to two fixed bone junctions, namely a stable wire fixation and an unstable wire fixation with a 10 mm wide diastasis mimicking a widely rupturing sternotomy. A generalized Linear Mixed Model with the lme function was used to determine the ability of the vibration transmittance device to differentiate mechanical settings in the sternotomy.

RESULTS

The transmitted vibration power was statistically significantly different between the intact chest and stable sternotomy closure, stable and unstable closure, as well as intact and unstable closure (t-values and p-values respectively: t = 6.87, p < 0.001; t = 7.41, p < 0.001; t = 14.3, p < 0.001). The decrease of vibration transmittance from intact to stable at all tested costal levels was 78%, from stable to unstable 58%, and from intact to unstable 91%. The vibration transmittance power was not statistically significantly different between the three tested costal levels (level 3 vs. level 2; level 4 vs. level 2; level 4 vs. level 3; t-values and p-values respectively t = - 0.36, p = 0.723; t = 0.35, p = 0.728; t = 0.71, p = 0.484).

CONCLUSIONS

Vibration transmittance analysis differentiates the intact sternum, wire fixation with exact apposition, and wire fixation with a gap. The gap detection capability is not dependent on the tested costal level. The method may prove useful in the early detection of sternal instability and warrants further exploration.

Для цитирования: Hautalahti Juha, Joutsen Atte, Goebeler Sirkka, Luukkaala Tiina, Khan Jahangir, Hyttinen Jari, Laurikka Jari. Vibration transmittance measures sternotomy stability - a preliminary study in human cadavers.. Journal of cardiothoracic surgery. 2019. DOI: 10.1186/s13019-018-0823-5.