MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · The Journal of invasive cardiology · 2008

Self-expanding platinum-coated nitinol devices for transcatheter closure of atrial septal defect: prevention of nickel release.

Lertsapcharoen Pornthep, Khongphatthanayothin Apichai, Srimahachota Suphot, Leelanukrom Ruenreong

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

A variety of nitinol-containing devices for transcatheter closure of atrial septal defects (ASD) has been widely used. However, there is concern about the release of nickel after nitinol device implantation. In this study, a platinum-coated nitinol device was braided from nanoplatinum-coated nitinol wires in order to prevent nickel release. The serum nickel levels before and after device implantation and the 1-year results were evaluated.

METHODS

Thirty-one patients, aged 4-59 years, and weighing 13.7-90.0 kg, underwent transcatheter closure. Blood samples for serum nickel levels were taken before, 1 day, 1 week, 1 month and 3 months after implantation.

RESULTS

Twenty-nine (93.6%) patients had a successful implantation. The mean ASD diameter was 19.7 +/- 4.8 mm (range 10-30 mm). Procedure-related complications included transient brachial plexus injury in 1 patient and transient dysrhythmia in 4 patients. All 29 patients had complete closure within 1 month after implantation. The mean serum nickel levels at baseline and at 1 day, 1 week, 1 month and 3 months after implantation were 0.65 +/- 0.28, 0.63 +/- 0.18, 0.67 +/- 0.34, 0.55 +/- 0.16, 0.52 +/- 0.14 ng/ml, respectively. There was no significant difference in serum nickel levels before and after implantation. There were no device-related complications at 1-year follow up.

CONCLUSIONS

Transcatheter ASD closure using a platinum-coated nitinol device can be performed safely and successfully with good outcomes. Nano-coating of platinum on nitinol wires can prevent nickel release following device implantation.

Для цитирования: Lertsapcharoen Pornthep, Khongphatthanayothin Apichai, Srimahachota Suphot, Leelanukrom Ruenreong. Self-expanding platinum-coated nitinol devices for transcatheter closure of atrial septal defect: prevention of nickel release.. The Journal of invasive cardiology. 2008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.