Percutaneous closure of hypertensive ductus arteriosus.
Аннотация
BACKGROUND
The Amplatzer duct occluder (ADO) has been used with success to close large patent ductus arteriosus (PDA), but some problems exist especially with hypertensive PDAs, such as incomplete closure, haemolysis, left pulmonary artery stenosis, obstruction of the descending aorta and progressive pulmonary vascular disease.
METHODS AND
RESULTS
We analysed a group of 168 patients with isolated PDA and pulmonary artery systolic pressure (PSAP) > or =50 mm Hg. Mean age was 10.3 +/- 14.3 years (median 3.9), PDA diameter was 6.4 +/- 2.9 mm (median 5.9), PASP was 63.5 +/- 16.2 mm Hg (median 60), Qp/Qs was 2.7 +/- 1.2 (median 2.5), total pulmonary resistance index (PRI) was 3.69 +/- 2.15 (median 3.35) and vascular PRI was 2.73 +/- 1.72 (median 2.37). We used ADOs in 145 (86.3%) cases, Amplatzer muscular ventricular septal defect occluders (AMVSDO) in 18 (10.7%), Amplatzer septal occluders (ASO) in three (1.8%) and the Gianturco-Grifka device in two (1.2%) cases. Device diameter was 106.3% +/- 51% higher than PDA diameter. PASP decreased after occlusion to 42.5 +/- 13.3 mm Hg (p<0.00001). Immediately after closure, no or trivial shunt was present in 123 (74.5%) cases. Immediate complications were device embolisation in five (3%) cases and descending aortic obstruction in one case. The overall success rate was 98.2%. Follow-up in 145 (86.3%) cases for 37.1 +/- 24 months (median 34.1) showed further decrease of the PASP to 30.1 +/- 7.7 mm Hg (p<0.0001).
CONCLUSIONS
Percutaneous treatment of hypertensive PDA is safe and effective. ADO works well for most cases, but sometimes other devices (MVSDO or ASO) have to be used. When cases are selected adequately, pulmonary pressures decrease immediately and continue to fall with time.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.