MedElevator Войти в исследования
← Поиск публикаций

DOI 10.6061/clinics/2020/e1764

Медицинская библиотека · Clinics (Sao Paulo, Brazil) · 2020

Accuracy of a portable breath meter test for the detection of halitosis in children and adolescents.

Guedes Carolina Cardoso, Bussadori Sandra Kalil, Garcia Ana Carolina Mota, Motta Lara Jansiski, Gomes Andréa Oliver, Weber Raimar, Amancio Olga Maria Silverio

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

This study aimed to determine the accuracy of the Breath-Alert™ portable breath meter (BA) for the detection of halitosis in children and adolescents, considering the organoleptic test (OT) as the gold standard in this assessment.

METHODS

A cross-sectional study was conducted on 150 children (aged 6-12 years). OT was performed by three independent examiners on a single occasion, obtaining three scores of 0-5 points on the Rosenberg's organoleptic scale. The median of the three evaluations for each child was used for analysis. BA was used according to the manufacturer's instructions, with breath odor scored from 0-5 points. Scores ≥2 on both tests were considered indicative of halitosis.

RESULTS

A total of 26 (17.3%) and 23 (15.3%) children were detected with halitosis on the OT and BA tests, respectively. The sensitivity and specificity of the BA scores for the detection of halitosis were 80.76% and 98.38%, respectively. The positive and negative predictive values for BA were 91.3% and 96.06%, respectively.

CONCLUSION

In the present study involving children, who require fast, practical examinations, BA proved to be an auxiliary tool to OT for the detection of halitosis in the practice of pediatric dentistry, demonstrating high sensitivity and specificity.

Для цитирования: Guedes Carolina Cardoso, Bussadori Sandra Kalil, Garcia Ana Carolina Mota, Motta Lara Jansiski, Gomes Andréa Oliver, Weber Raimar, Amancio Olga Maria Silverio. Accuracy of a portable breath meter test for the detection of halitosis in children and adolescents.. Clinics (Sao Paulo, Brazil). 2020. DOI: 10.6061/clinics/2020/e1764.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.