MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ijcard.2025.133068

Медицинская библиотека · International journal of cardiology · 2025

Remote dielectric sensing to detect pulmonary congestion in acute dyspnoeic patients: Reproducibility and the effect of pulmonary comorbidities.

Olesen Anne Sophie Overgaard, Miger Kristina, Sajadieh Ahmad, Abild-Nielsen Annemette Geilager, Pedersen Lars, Schultz Hans Henrik Lawaetz, Grand Johannes, Thune Jens Jakob, Nielsen Olav Wendelboe

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Remote Dielectric Sensing (ReDS) is a fast and non-invasive method that estimates lung fluid. We previously described moderate accuracies for ReDS to detect acute heart failure in consecutive patients. We hypothesise that unprecise ReDS values may stem from concomitant pulmonary diseases.

PURPOSE: To examine the ReDS reproducibility and the effect of pulmonary comorbidities on ReDS values in acute dyspnoeic patients.

METHODS

This prospective observational study included 97 consecutive patients ≥50 years with acute dyspnoea. Upon admission, patients underwent low-dose chest computed tomography (CT), echocardiography and ReDS examination. ReDS is by default performed on the right hemithorax in sitting position. For reproducibility comparisons, we conducted additional ReDS measurements two centimetres above and below the default placement, and in sitting and supine position. Two blinded radiologists evaluated the CT scans for pulmonary congestion and pulmonary diseases.

RESULTS

Comparing three ReDS measurements on the right hemithorax revealed coefficients of variations of 9.6 %, 8.2 %, and 8.3 %. For sitting versus supine comparison, the coefficient of variation was 9.5 % for the default ReDS placement. Patients with CT-verified pulmonary congestion had a coefficient of variation of 5.9 % in sitting versus supine comparison, while those without had 10.3 %. In multivariable regression, lower ReDS values were observed in patients with pneumonia (-1.81, p = 0.215, N = 51), emphysema (-5.44, p = 0.001, N = 26), and higher in fibrosis (5.58, p = 0.032, N = 8) and congestion (5.79, p = 0.002, N = 17), compared to those without.

CONCLUSION

ReDS values of lung fluid content and reproducibility were affected by pulmonary diseases. ReDS showed consistent reproducibility for patients with CT-verified pulmonary congestion.

Для цитирования: Olesen Anne Sophie Overgaard, Miger Kristina, Sajadieh Ahmad, Abild-Nielsen Annemette Geilager, Pedersen Lars, Schultz Hans Henrik Lawaetz, Grand Johannes, Thune Jens Jakob, Nielsen Olav Wendelboe. Remote dielectric sensing to detect pulmonary congestion in acute dyspnoeic patients: Reproducibility and the effect of pulmonary comorbidities.. International journal of cardiology. 2025. DOI: 10.1016/j.ijcard.2025.133068.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.