Zhou Yong, Qin Yongwen, Zhao Xianxian, Lang Xilong, Zhu Ni, Bai Yuan, Zheng Xing, Guo Zhifu, Wu Hong
Аннотация
BACKGROUND: Complete atrioventricular block (cAVB) is considered the most serious adverse event after transcatheter closure of ventricular septal defect (VSD).
OBJECTIVES: To evaluate the incidence of postprocedure cAVB and its relationship with different waist lengths of transcatheter occluders.
METHODS: This is a retrospective study of patients with VSD who had been treated with transcatheter closure at the Shanghai Changhai Hospital from December 2001 to December 2010. Patients were treated with short-waist (n = 234) or long-waist occluders (n = 571). All patients were monitored by electrocardiogram (ECG) within 7 days post procedure.
RESULTS: Nine patients (3.8%) treated with short-waist occluders suffered from cAVB and 3 patients (1.3%) required permanent pacemaker implantation, compared with 4 patients (0.7%) and 1 patient (0.2%), respectively, in the long-waist occluder group (P<.001). There was a significantly higher incidence of postprocedure incomplete right bundle branch block (IRBBB) in patients treated with short-waist occluders compared with patients treated with long-waist occluders. There were no significant differences in other procedure-related complications between the two groups.
CONCLUSIONS: Long-waist VSD occluders are beneficial in the prevention of cAVB and the need for pacemaker implantation after transcatheter closure of VSD.
OBJECTIVES: To evaluate the incidence of postprocedure cAVB and its relationship with different waist lengths of transcatheter occluders.
METHODS: This is a retrospective study of patients with VSD who had been treated with transcatheter closure at the Shanghai Changhai Hospital from December 2001 to December 2010. Patients were treated with short-waist (n = 234) or long-waist occluders (n = 571). All patients were monitored by electrocardiogram (ECG) within 7 days post procedure.
RESULTS: Nine patients (3.8%) treated with short-waist occluders suffered from cAVB and 3 patients (1.3%) required permanent pacemaker implantation, compared with 4 patients (0.7%) and 1 patient (0.2%), respectively, in the long-waist occluder group (P<.001). There was a significantly higher incidence of postprocedure incomplete right bundle branch block (IRBBB) in patients treated with short-waist occluders compared with patients treated with long-waist occluders. There were no significant differences in other procedure-related complications between the two groups.
CONCLUSIONS: Long-waist VSD occluders are beneficial in the prevention of cAVB and the need for pacemaker implantation after transcatheter closure of VSD.
Предметные рубрики MeSH
AdolescentAdultAtrioventricular BlockCardiac CatheterizationCardiac Surgical ProceduresChildChild, PreschoolElectrocardiographyEquipment DesignFemaleFollow-Up StudiesHeart Septal Defects, VentricularHumansMaleMiddle AgedPostoperative ComplicationsRetrospective StudiesSeptal Occluder DeviceTreatment OutcomeYoung Adult