Photoplethysmographic assessment of pulse transit time correlates with echocardiographic measurement of stroke volume in preterm infants with patent ductus arteriosus.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Amirtharaj Cynthia R, Palmeri Lynn C, Gradwohl Gideon, Adar Yair, Nitzan Meir, Gruber Dorota, Blaufox Andrew, Meyer David B, Kwon Elena N, Koppel Robert
Аннотация
OBJECTIVE: We aimed to correlate photoplethysmographic parameters with stroke volume in infants with PDA. Photoplethysmography constitutes the optical signal in pulse oximetry.
STUDY DESIGN: Stroke volume was determined echocardiographically. Pulse transit time, right hand to foot arrival time difference, and relative amplitude were measured from pulse oximeter and ECG waveforms. Photoplethysmographic parameters before and after PDA closure were compared with stroke volume.
RESULTS: After PDA closure, pulse transit time to the hand and to the foot were prolonged (54.7 ± 6.7 vs 65.5 ± 9.8 ms, p < 0.001, 82.5 ± 12.8 vs 88.6 ± 10.6 ms, p = 0.03), arrival time difference decreased (27.7 ± 7.6 vs 23.1 ± 5.6 ms, p = 0.021), and relative amplitude decreased (from 2.1 ± 0.7% to 1.5 ± 0.5%, p = 0.003). The time-based photoplethysmographic parameters correlated with stroke volume.
CONCLUSIONS: Photoplethysmographic waveform parameters are significantly different before and after PDA closure and the time-based parameters correlate well with stroke volume. Monitoring pulse transit time may assist in evaluation for spontaneous PDA closure or response to therapy.
STUDY DESIGN: Stroke volume was determined echocardiographically. Pulse transit time, right hand to foot arrival time difference, and relative amplitude were measured from pulse oximeter and ECG waveforms. Photoplethysmographic parameters before and after PDA closure were compared with stroke volume.
RESULTS: After PDA closure, pulse transit time to the hand and to the foot were prolonged (54.7 ± 6.7 vs 65.5 ± 9.8 ms, p < 0.001, 82.5 ± 12.8 vs 88.6 ± 10.6 ms, p = 0.03), arrival time difference decreased (27.7 ± 7.6 vs 23.1 ± 5.6 ms, p = 0.021), and relative amplitude decreased (from 2.1 ± 0.7% to 1.5 ± 0.5%, p = 0.003). The time-based photoplethysmographic parameters correlated with stroke volume.
CONCLUSIONS: Photoplethysmographic waveform parameters are significantly different before and after PDA closure and the time-based parameters correlate well with stroke volume. Monitoring pulse transit time may assist in evaluation for spontaneous PDA closure or response to therapy.