DOI 10.1016/j.carpath.2016.08.007

Медицинская библиотека · Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology · 2016

Peristrut microhemorrhages: a possible cause of in-stent neoatherosclerosis?

Terzian Zaven, Gasser T Christian, Blackwell Francis, Hyafil Fabien, Louedec Liliane, Deschildre Catherine, Ghodbane Walid, Dorent Richard, Nicoletti Antonino, Morvan Marion, Nejjari Mohammed, Feldman Laurent, Pavon-Djavid Graciela, Michel Jean-Baptiste

СтатьяПеревод в очереди

Аннотация

BACKGROUND

In-stent neoatherosclerosis is characterized by the delayed appearance of markers of atheroma in the subintima, but the pathophysiology underlying this new disease entity remains unclear.

METHODS AND

RESULTS

We collected 20 human coronary artery stents by removal from explanted hearts. The mean duration of stent implantation was 34 months. In all samples, neoatherosclerosis was detected, particularly in peristrut areas. It consisted of foam cells and cholesterol clefts, with or without calcification, associated with neovascularization. Iron and glycophorin-A were present in peristrut areas, as well as autofluorescent ceroids. Moreover, in response to neoatherosclerosis, tertiary lymphoid organs (tissue lymphoid clusters) often developed in the adventitia. Some of these features could be reproduced in an experimental carotid stenting model in rabbits fed a high-cholesterol diet. Foam cells were present in all samples, and peristrut red blood cells (RBCs) were also detected, as shown by iron deposits and Bandeiraea simplicifiola isolectin-B4 staining of RBC membranes. Finally, in silico models were used to evaluate the compliance mismatch between the rigid struts and the distensible arterial wall using finite element analysis. They show that stenting approximately doubles the local von Mises stress in the intimal layer.

CONCLUSIONS

We show here that stent implantation both in human and in rabbit arteries is characterized by local peristrut microhemorrhages and finally by both cholesterol accumulation and oxidation, triggering together in-stent neoatherosclerosis. Our data indicate that these processes are likely initiated by an increased mechanical stress due to the compliance mismatch between the rigid stent and the soft wall.

Для цитирования: Terzian Zaven, Gasser T Christian, Blackwell Francis, Hyafil Fabien, Louedec Liliane, Deschildre Catherine, Ghodbane Walid, Dorent Richard, Nicoletti Antonino, Morvan Marion, Nejjari Mohammed, Feldman Laurent, Pavon-Djavid Graciela, Michel Jean-Baptiste. Peristrut microhemorrhages: a possible cause of in-stent neoatherosclerosis?. Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology. 2016. DOI: 10.1016/j.carpath.2016.08.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.