DOI 10.1097/prs.0000000000008821
Elevated Intracranial Pressure in Patients with Craniosynostosis by Optical Coherence Tomography.
Аннотация
BACKGROUND
Craniosynostosis may lead to elevated intracranial pressure, which may be implicated with impaired neurocognitive development. However, accurately measuring intracranial pressure is challenging, and patterns in craniosynostosis patients are poorly characterized. Spectral-domain optical coherence tomography may enable noninvasive assessment of intracranial pressure in pediatric patients with craniosynostosis.
METHODS
Pediatric patients with craniosynostosis undergoing surgical intervention between 2014 and 2019 prospectively underwent optical coherence tomographic evaluation. Intracranial pressure was directly measured intraoperatively in a subset of cases. Optical coherence tomographic parameters were compared to directly measured intracranial pressure and used for pattern assessment.
RESULTS
Optical coherence tomography was performed in 158 subjects, among which 42 underwent direct intracranial pressure measurement during an initial cranial procedure. Maximal retinal nerve fiber layer thickness, maximal retinal thickness, and maximal anterior projection optical coherence tomographic parameters were positively correlated with intracranial pressure (p ≤ 0.001), with all parameters showing significantly higher values in patients with intracranial pressure thresholds of 15 mmHg (p < 0.001) and 20 mmHg (p ≤ 0.007). Patients with maximal retinal nerve fiber layer thickness and maximal anterior projection exceeding set thresholds in optical coherence tomography of either eye demonstrated 77.3 percent sensitivity and 95.0 percent specificity for detecting intracranial pressure above 15 mmHg, and 90.0 percent sensitivity and 81.3 percent specificity for detecting intracranial pressure above 20 mmHg. Patients with associated syndromes or multiple suture involvement and patients aged 9 months or older were significantly more likely to have elevated intracranial pressure above 15 mmHg (p ≤ 0.030) and above 20 mmHg (p ≤ 0.035).
CONCLUSIONS
Spectral-domain optical coherence tomography can noninvasively detect elevated intracranial pressure in patients with craniosynostosis with reliable sensitivity and specificity. This technology may help guide decisions about the appropriate type and timing of surgical treatment.
CLINICAL QUESTION/LEVEL OF EVIDENCE: Diagnostic, I.