Smartphone Magnetometers for Assessing Programmable Shunt Valve Settings: A Proof-of-Concept Study.

McCoy Carlie, Patel Smruti K, Skoch Jesse

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND

OBJECTIVES

Programmable shunt valves (PSVs) used to treat hydrocephalus require specialized valve-specific devices that determine their setting. However, these devices are typically only accessible in major medical centers. This study explores the feasibility of using smartphone magnetometers to assess PSV settings.

METHODS

We investigated 2 PSVs, the Codman Certas Plus and Medtronic Strata II, using 3 different smartphones. Free open-source software was used to measure magnetic flux densities of the shunts, and PSV-specific algorithms were developed to interpret these measurements. All measurements were performed ex vivo.

RESULTS

We demonstrated that this method could accurately determine PSV settings with 100% accuracy in ex vivo blinded trials. The technique was effective even with a scalp analog simulating different skin thicknesses.

CONCLUSION

This proof-of-concept study suggests that smartphones can be used for PSV assessment. However, further development and validation are needed to refine this method for clinical use. This technology could enhance PSV management accessibility, allowing for remote or at-home evaluations by trained individuals or caregivers, thus improving patient care in underserved areas.

Для цитирования: McCoy Carlie, Patel Smruti K, Skoch Jesse. Smartphone Magnetometers for Assessing Programmable Shunt Valve Settings: A Proof-of-Concept Study.. Neurosurgery. 2024. DOI: 10.1227/neu.0000000000003221.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.