Becher Tobias H, Miedema Martijn, Kallio Merja, Papadouri Thalia, Karaoli Christina, Sophocleous Louiza, Rahtu Marika, van Leuteren Ruud W, Waldmann Andreas D, Strodthoff Claas, Yerworth Rebecca, Dupré Antoine, Benissa Mohamed-Rida, Nordebo Sven, Khodadad Davood, Bayford Richard, Vliegenthart Roseanne, Rimensberger Peter C, van Kaam Anton H, Frerichs Inéz

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Rationale: Electrical impedance tomography (EIT) allows instantaneous and continuous visualization of regional ventilation and changes in end-expiratory lung volume at the bedside. There is particular interest in using EIT for monitoring in critically ill neonates and young children with respiratory failure. Previous studies have focused only on short-term monitoring in small populations. The feasibility and safety of prolonged monitoring with EIT in neonates and young children have not been demonstrated yet.

Objectives

To evaluate the feasibility and safety of long-term EIT monitoring in a routine clinical setting and to describe changes in ventilation distribution and homogeneity over time and with positioning in a multicenter cohort of neonates and young children with respiratory failure.

Methods

At four European University hospitals, we conducted an observational study (NCT02962505) on 200 patients with postmenstrual ages (PMA) between 25 weeks and 36 months, at risk for or suffering from respiratory failure. Continuous EIT data were obtained using a novel textile 32-electrode interface and recorded at 48 images/s for up to 72 hours. Clinicians were blinded to EIT images during the recording. EIT parameters and the effects of body position on ventilation distribution were analyzed offline.

Results

The average duration of EIT measurements was 53 ± 20 hours. Skin contact impedance was sufficient to allow image reconstruction for valid ventilation analysis during a median of 92% (interquartile range, 77-98%) of examination time. EIT examinations were well tolerated, with minor skin irritations (temporary redness or imprint) occurring in 10% of patients and no moderate or severe adverse events. Higher ventilation amplitude was found in the dorsal and right lung areas when compared with the ventral and left regions, respectively. Prone positioning resulted in an increase in the ventilation-related EIT signal in the dorsal hemithorax, indicating increased ventilation of the dorsal lung areas. Lateral positioning led to a redistribution of ventilation toward the dependent lung in preterm infants and to the nondependent lung in patients with PMA > 37 weeks.

Conclusions

EIT allows continuous long-term monitoring of regional lung function in neonates and young children for up to 72 hours with minimal adverse effects. Our study confirmed the presence of posture-dependent changes in ventilation distribution and their dependency on PMA in a large patient cohort. Clinical trial registered with www.clinicaltrials.gov (NCT02962505).

Для цитирования: Becher Tobias H, Miedema Martijn, Kallio Merja, Papadouri Thalia, Karaoli Christina, Sophocleous Louiza, Rahtu Marika, van Leuteren Ruud W, Waldmann Andreas D, Strodthoff Claas, Yerworth Rebecca, Dupré Antoine, Benissa Mohamed-Rida, Nordebo Sven, Khodadad Davood, Bayford Richard, Vliegenthart Roseanne, Rimensberger Peter C, van Kaam Anton H, Frerichs Inéz. Prolonged Continuous Monitoring of Regional Lung Function in Infants with Respiratory Failure.. Annals of the American Thoracic Society. 2022. DOI: 10.1513/annalsats.202005-562oc.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.