A single-arm, open-label, multicenter, and comparative study of the ANNE sleep system vs polysomnography to diagnose obstructive sleep apnea.
Аннотация
STUDY
OBJECTIVES
Evaluate per-patient diagnostic performance of a wireless dual-sensor system (ANNE sleep) compared with reference standard polysomnography (PSG) for the diagnosis of moderate and severe obstructive sleep apnea (OSA) with a minimum prespecified threshold of 80% for both sensitivity and specificity.
METHODS
A multicenter clinical trial was conducted to evaluate ANNE sleep vs PSG to diagnose moderate and severe OSA in individuals 22 years or older. For each testing approach, apnea-hypopnea index (AHI) was manually scored and averaged by 3 registered sleep technologists blinded to the other system. Average variations > 15% were adjudicated by a sleep medicine physician.
RESULTS
In a total of n = 225 participants (mean age 53 years, range 22-88 years), PSG diagnosed 30% (n = 68) of participants with moderate or severe OSA (AHI ≥ 15 events/h) compared to 29% (n = 65) diagnosed by ANNE sleep (P = .55). The sensitivity and specificity for ANNE sleep were 90% (95% confidence interval: 80-96%) and 98% (95% confidence interval: 94-99%), respectively. Strong correlation was shown in terms of final AHI (r = .93), with an average AHI bias of 0.5 (95% limits of agreement: -12.8 to 11.8). The majority of users noted comfort with using the ANNE sleep in the home setting. No adverse events were noted.
CONCLUSIONS
Using PSG as the gold standard, ANNE sleep demonstrated high sensitivity and specificity for the diagnosis of moderate or severe OSA.
CLINICAL TRIAL REGISTRATION: Registry: ClinicalTrials.gov; Name: Comparative Study of the ANNE™ One System to Diagnose Obstructive Sleep Apnea; URL: https://clinicaltrials.gov/ct2/show/NCT04643782; Identifier: NCT04643782.
CITATION: Davies C, Lee JY, Walter J et al. A single-arm, open-label, multicenter, and comparative study of the ANNE sleep system vs polysomnography to diagnose obstructive sleep apnea. J Clin Sleep Med. 2022;18(12):2703-2712.