[Phantom studies of ultrasound equipment for quality improvement in breast diagnosis].
Аннотация
AIM: According to the German guidelines for quality control of ultrasonic equipment, the following conditions are required for breast ultrasound: A transducer frequency between 5-7.5 MHz and a minimum field of view of 5 cm. Satisfactory images must be obtained in a depth between 0.5 and 4 cm with a wide tolerance of the focal zones. This allows the use of poor quality equipment which does not produce satisfactory image quality and it excludes a number of high frequency and high resolution transducers with a field of view below 5 cm. This study with a test phantom was performed to define image quality objectively.
METHOD: Sixteen ultrasound instruments in different price categories were used to perform standardized examinations on a breast phantom model 550 (ATS Laboratories, Bridgeport, USA). Contrast and spatial resolution in different penetration depths were investigated on cyst phantoms from 1-4 mm diameter and wire targets with defined distances between 0.5-3 mm 4 investigations reported the images.
RESULTS
A positive correlation was seen between price category and image quality.
CONCLUSION
This study demonstrates that transducer frequency and image geometry do not allow sufficient quality control. An improvement of ultrasound diagnosis is only possible if equipment guidelines are based on standard examinations with test phantoms.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.