The computer-aided parallel external fixator for complex lower limb deformity correction.

Wei Mengting, Chen Jianwen, Guo Yue, Sun Hao

СтатьяПеревод в очереди

Аннотация

PURPOSE: Since parameters of the parallel external fixator are difficult to measure and calculate in real applications, this study developed computer software that can help the doctor measure parameters using digital technology and generate an electronic prescription for deformity correction.

METHODS

According to Paley's deformity measurement method, we provided digital measurement techniques. In addition, we proposed an deformity correction algorithm to calculate the elongations of the six struts and developed a electronic prescription software. At the same time, a three-dimensional simulation of the parallel external fixator and deformed fragment was made using virtual reality modeling language technology. From 2013 to 2015, fifteen patients with complex lower limb deformity were treated with parallel external fixators and the self-developed computer software. All of the cases had unilateral limb deformity. The deformities were caused by old osteomyelitis in nine cases and traumatic sequelae in six cases. A doctor measured the related angulation, displacement and rotation on postoperative radiographs using the digital measurement techniques. Measurement data were input into the electronic prescription software to calculate the daily adjustment elongations of the struts. Daily strut adjustments were conducted according to the data calculated. The frame was removed when expected results were achieved. Patients lived independently during the adjustment.

RESULTS

The mean follow-up was 15 months (range 10-22 months). The duration of frame fixation from the time of application to the time of removal averaged 8.4 months (range 2.5-13.1 months). All patients were satisfied with the corrected limb alignment. No cases of wound infections or complications occurred.

CONCLUSIONS

Using the computer-aided parallel external fixator for the correction of lower limb deformities can achieve satisfactory outcomes. The correction process can be simplified and is precise and digitized, which will greatly improve the treatment in a clinical application.

Для цитирования: Wei Mengting, Chen Jianwen, Guo Yue, Sun Hao. The computer-aided parallel external fixator for complex lower limb deformity correction.. International journal of computer assisted radiology and surgery. 2017. DOI: 10.1007/s11548-017-1654-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.