DOI 10.3760/cma.j.cn121094-20230103-000001
[Determination of 4-methyl-2-pentanol in workplace air by solvent desorption-gas chromatography].
Аннотация
Objective
To establish a method for the determination of 4-methyl-2-pentanol in the air of workplace by gas chromatography.
Methods
In January 2022, 4-methyl-2-pentanol in the air of workplace was collected by activated carbontube, eluted with dichloromethane-methanol (95∶5, V/V), separated by capillary column and determined by gas chromatogram.
Results
The limit of detection for 4-methyl-2-pentanol was 0.04 μg/ml. The linear range of 4-methyl-2-pentanol was 0.16-1616.60 μg/ml, with the regression equation of y=1.94x-5.48, and the coefficient correlation was 0.99958, and the minimum detection concentration was 0.03 mg/m(3) (collected sample volume was 1.50 L). The within-run precisions were 1.08%-1.75% and the between-run precisions were 1.41%-2.52%. The desorption rates were 95.15%-99.91%. The samples could be stored at least 3 days at room temperature and 7 days at 4 ℃ without significant loss.
Conclusion
The method has the advantages of good precision, high sensitivity and simple operation. It is suitable for the determination of 4-methyl-2-pentanol in the air of workplace.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.