MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.wneu.2019.06.095

Медицинская библиотека · World neurosurgery · 2019

Unexpectedly Smaller Artifacts of 3.0-T Magnetic Resonance Imaging than 1.5 T: Recommendation of 3.0-T Scanners for Patients with Magnet-Resistant Adjustable Ventriculoperitoneal Shunt Devices.

Amano Yuki, Kuroda Naoto, Uchida Daiki, Sakakura Yuya, Nakatogawa Hirokazu, Ando Naoto, Nakayama Teiji, Sato Haruhiko, Masui Takayuki, Sameshima Tetsuro, Tanaka Tokutaro

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Magnetic resonance imaging (MRI) artifacts of adjustable shunt devices are thought to be similar to metal clip artifacts, in that they are larger with higher field strength scanners. We have published several reports about the artifacts of new MRI-resistant adjustable shunt devices, and we found a case in which a 3.0-T scanner showed smaller artifacts than the 1.5-T scanner. We aimed to clarify whether this claim is true or not.

METHODS

Under permission of our institutional Ethical Committee, 2 volunteers underwent imaging studies using 3.0-T and 1.5-T scanners from GE, Siemens, and Philips. Four MRI-resistant adjustable shunt devices-proGAV2.0 (Miethke), Codman Certas Plus (Johnson & Johnson), Polaris (Sophysa), and Strata MR valve (Medtronic)-were fixed on the left temporal scalp. Routine MRI images, including T1-and T2-weighted imaging, fluid-attenuated inversion recovery, diffusion-weighted imaging (DWI), and magnetic resonance angiography (MRA), were obtained. We also compared artifacts between a 3.0-T scanner and a-1.5 T scanner in 4 patients.

RESULTS

The 3.0 T-scanners showed smaller artifacts than the 1.5-T scanners on DWI and MRA images for all shunt devices and scanners. In the other sequences, the results depended on the MRI scanner manufacturer; however, the GE 3.0-T scanner showed smaller artifacts in every sequence. This was also true in the 4 clinical cases.

CONCLUSIONS

A 3.0-T scanner is recommended over a 1.5-T scanner for patients with MRI-resistant adjustable shunt devices in the diagnosis of acute ischemic condition or when using GE scanners.

Для цитирования: Amano Yuki, Kuroda Naoto, Uchida Daiki, Sakakura Yuya, Nakatogawa Hirokazu, Ando Naoto, Nakayama Teiji, Sato Haruhiko, Masui Takayuki, Sameshima Tetsuro, Tanaka Tokutaro. Unexpectedly Smaller Artifacts of 3.0-T Magnetic Resonance Imaging than 1.5 T: Recommendation of 3.0-T Scanners for Patients with Magnet-Resistant Adjustable Ventriculoperitoneal Shunt Devices.. World neurosurgery. 2019. DOI: 10.1016/j.wneu.2019.06.095.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.