MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s0161-6420(00)00603-5

Медицинская библиотека · Ophthalmology · 2001

Clinical evaluation and risk factors of time to failure of Ahmed Glaucoma Valve implant in pediatric patients.

Djodeyre M R, Peralta Calvo J, Abelairas Gomez J

Клиническое исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the clinical results and the risk factors related to time to failure of the Ahmed Glaucoma Valve implant (New World Medical, Inc., Rancho Cucamonga, CA), an aqueous shunting device with a unidirectional valve mechanism, in pediatric patients.

DESIGN

Retrospective, noncomparative, interventional case series.

PARTICIPANTS

We studied 35 Ahmed Glaucoma Valve implants in 35 eyes of 29 patients less than 15 years of age (median, 2.4 years; range, 0-14.8 years).

METHODS

Thirty-five Ahmed Glaucoma Valve implants were placed in eyes of pediatric patients with refractory glaucoma between December 1994 and March 1999.

MAIN OUTCOME MEASURES: Time after surgery without failure, intraocular pressure reduction, risk factors related to time to failure, and postoperative complications were evaluated.

RESULTS

The cumulative probabilities of success according to Kaplan-Meier analysis 12 and 24 months after implantation were 70.1% +/- 8.5% and 63.7% +/- 9.9%, respectively. Intraocular pressure was reduced from a preoperative mean of 28.8 +/- 4.5 mmHg to 18.1 +/- 2.4 mmHg by postoperative month 18. Shallow anterior chambers of different grades developed in nine eyes (25.7%) on the first postoperative day; this was associated with choroidal detachment in four eyes. Nine eyes (25.7%) experienced malposition, causing failure in six eyes. In three other eyes (8.5%), the average intraocular pressure at the last two follow-up examinations was more than 22 mmHg. Congenital glaucoma and surgical experience were significantly associated with time to failure.

CONCLUSIONS

The 12- and 24-month success rates of the Ahmed Glaucoma Valve implant in our study confirm its viability in the management of refractory infantile glaucoma. The complications related to overfiltration during the immediate postoperative period appear to be less severe than those associated with other drainage devices and resolve spontaneously. Diagnosis, number of previous glaucoma procedures, and the surgeon's experience seem to be related to the survival of the Ahmed Glaucoma Valve implant.

Для цитирования: Djodeyre M R, Peralta Calvo J, Abelairas Gomez J. Clinical evaluation and risk factors of time to failure of Ahmed Glaucoma Valve implant in pediatric patients.. Ophthalmology. 2001. DOI: 10.1016/s0161-6420(00)00603-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.