Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2002

Efficiency of a video-autorefractometer used as a screening device for amblyogenic factors.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Schimitzek Thilo, Haase Wolfgang

Аннотация

BACKGROUND: High ametropia is known as a risk factor in developing amblyopia. The aim of the study was to evaluate the efficiency of a commercially available infrared videorefractometer (PowerRefractor) in examining children, and to evaluate its effectiveness in detecting amblyogenic factors of the refraction. The influence of cycloplegia was of interest.

METHODS: A total of 255 eyes (-28.25 D to +7.88 D spherical equivalent) of 128 patients (1-81 years) were examined by cycloplegic retinoscopy. All of them were measured with the PowerRefractor under cycloplegia. Fifty-four of these patients (108 eyes) were additionally measured without cycloplegia. The technique of the device is based on the eccentric photorefraction method. Both eyes can be measured simultaneously from a distance of 1 m. A video-analysis of each eye is computed in three meridians. The subject can be observed in real time on a screen, and the refraction is immediately available.

RESULTS: Measurement of refraction was possible at all ages. More effort was required with young children who could not cooperate well with the procedure of conventional table-top autorefractometers. The mean sensitivity in detection of any amblyogenic ametropia was 80% with and without cycloplegia. The specificity was 61% with cycloplegia and 84% without. The rate of false-negative results in the detection of anisometropia, hyperopia and astigmatism ranged from 10% to 24%. In particular, high astigmatism and hyperopia remained undiscovered.

CONCLUSIONS: The reliability of the photorefractor which was the subject of the evaluation in detecting amblyogenic ametropia was comparable to that of other devices described previously in the literature. The method is suitable for use in children, but anisometropia, high astigmatic and hyperopic ametropia remained undiscovered in several cases. Commercial interest: none.

Предметные рубрики MeSH