Rapid communications in mass spectrometry : RCM · 2012

A simplified method for busulfan monitoring using dried blood spot in combination with liquid chromatography/tandem mass spectrometry.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ansari Marc, Uppugunduri Chakradhara Rao S, Déglon Julien, Théorêt Yves, Versace François, Gumy-Pause Fabienne, Ozsahin Hulya, Dayer Pierre, Desmules Jules, Daali Youssef

Аннотация

RATIONALE: Busulfan (Bu) is an important component of the myeloablative conditioning regimen prior to hematopoietic stem cell transplantation (HSCT) especially in children. Intravenously administered Bu exhibits a therapeutic window phenomenon requiring therapeutic drug monitoring. Analytical methods developed for Bu routine monitoring were aimed at using low volumes of biological fluids and development of simple procedures to facilitate the dosage adjustment. In this report, we describe a simple, rapid method for Bu measurement using dried blood spots (DBS) from only 5 μL of whole blood.

METHODS: Bu extracted from DBS with methanol was measured by high-performance liquid chromatography with electrospray ionization and tandem mass spectrometry in multiple reaction monitoring mode using D8-Bu as an internal standard. The method was in-house validated evaluating trueness, repeatability, within-laboratory reproducibility, specificity and the lower limit of quantification (LLOQ).

RESULTS: The method was linear in the calibration range of 100-2000 ng mL(-1) (r(2)>0.99) encompassing the therapeutic concentrations of Bu. A good trueness (<14%), precision (<10%), and recovery (100%) were observed during validation of the method with quality controls of 300, 600 and 1400 ng mL(-1). The LLOQ was determined as 50 ng mL(-1) and no matrix or carryover effects were observed. The validated method was applied to measure Bu levels in four children receiving infusion of Bu prior to HSCT. A good correlation was observed between the Bu levels measured by DBS and dried plasma spot (DPS) (r(2) =0.96) and between DPS and the GC/MS method (r(2) =0.92). Bu was found to be stable in DBS up to 6 h at room temperature and for 24 h at 4 °C.

CONCLUSIONS: The new DBS method facilitates earlier dosage adjustment during Bu therapy by its specific and simple procedure using 5 μL of whole blood.