Variation in the Phototherapy Practices and Irradiance of Devices in a Major Metropolitan Area.

Borden Amy R, Satrom Katie M, Wratkowski Paul, George Thomas N, Adkisson Constance A, Vreman Hendrik J, Johnson Austin P, Nichols Kent J, Slusher Tina M

Оценочное исследованиеСтатьяМногоцентровое исследованиеПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

BACKGROUND

Phototherapy (PT) is widely used to prevent and treat severe hyperbilirubinemia and its associated risks for both acute and chronic bilirubin encephalopathy. Intensive PT, recommended for inpatient treatment of hyperbilirubinemia in term and near-term infants, is defined as having a spectral irradiance of ≥30 μW/cm2/nm.

OBJECTIVES

We aimed to assess local PT practices by measuring the irradiance of PT devices in local neonatal intensive care units and newborn nurseries.

METHODS

The irradiance footprint, including maximum irradiance at the center of the footprint, of 39 PT devices in 7 area hospitals was measured according to current practice in these facilities.

RESULTS

The mean ± SD (range) footprint irradiance was 20.7 ± 5.8 (8.8-29.4) μW/cm2/nm. The mean ± SD maximum irradiance at the footprint center for all devices at a mean clinically used treatment distance of 33.1 ± 9.3 (25.5-60.0) cm was 27.8 ± 7.0 (14.7-42.0) μW/cm2/nm. Sixty-two percent of the devices did not meet the minimum recommended spectral irradiance for intensive PT. For the sites without irradiance-based protocols, the maximum irradiance of the devices (n = 33) at the treatment distances was 25.8 ± 6.1 μW/cm2/nm.

CONCLUSIONS

Despite established PT guidelines, local protocols and practices vary. Based on an assessment of 7 local hospitals, intensive PT was suboptimal for 62% of devices. Straightforward changes, such as decreasing the distance between an infant and the light source and establishing a consistent irradiance-based protocol, could substantially improve the quality of the intervention.

Для цитирования: Borden Amy R, Satrom Katie M, Wratkowski Paul, George Thomas N, Adkisson Constance A, Vreman Hendrik J, Johnson Austin P, Nichols Kent J, Slusher Tina M. Variation in the Phototherapy Practices and Irradiance of Devices in a Major Metropolitan Area.. Neonatology. 2018. DOI: 10.1159/000485369.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.