A rare postoperative complication of anterior lower thoracic instrumentation: diaphragmatic laceration with hemothorax.
Аннотация
PURPOSE: To highlight the importance of preventing visceral injury due to prominent anterior implants at the spinal column.
METHODS
A 52-year-old man with cord compression was treated with a T9/10 discectomy and instrumented fusion via a right thoracotomy and trans-pleural approach. Postoperatively, he had improved lower limb numbness. However, after a bout of coughing, there was sudden increase in chest drain output, hemodynamic instability and massive fluid collection in the right chest cavity.
RESULTS
Emergency re-exploration of the thoracotomy was performed and a 6 cm laceration of the right postero-medial diaphragm was identified as the bleeding source and was found to be in close proximity with the locking nut of the anterior implants. The laceration was repaired and a soft synthetic patch was used to cover the implants. Postoperatively, the hemothorax resolved and the patient recovered from the neurological deficit.
CONCLUSIONS
Prevention of diaphragmatic injury can be performed using lower profile and less sharp-edged implants. Implant coverage with a soft synthetic material is necessary if unable to perform direct repair of the parietal pleura over the implants.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.