MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvir.2018.06.015

Медицинская библиотека · Journal of vascular and interventional radiology : JVIR · 2018

Diagnostic Accuracy of Noncontrast MR Angiography Protocols at 3T for the Detection and Characterization of Lower Extremity Peripheral Arterial Disease.

Hanrahan Christopher J, Lindley Marc D, Mueller Michelle, Kim Daniel, Sommers Daniel, Morrell Glen, Redd Andrew, Carlston Kristi, Lee Vivian S

СтатьяПоддержка NIH: внешняя программаПоддержка правительства СШАПеревод в очереди

Аннотация

PURPOSE: To compare the diagnostic accuracy of established non-gadolinium (Gd)-enhanced magnetic resonance (MR) angiography protocols with Gd-enhanced MR angiography at 3T for evaluating lower extremity peripheral arterial disease (PAD).

MATERIALS AND

METHODS

From February 2014 to 2015, 20 patients with PAD and intermittent claudication (16 men; age range, 51-76 y; Fontaine stage II) underwent 3-station (abdominopelvic, thigh, and calf) non-Gd MR angiography and bolus-chase Gd MR angiography protocols performed at 3T (Siemens Tim Trio), including quiescent-interval single-shot (QISS) MR angiography for all 3 stations and a combination of quadruple inversion recovery (QIR) MR angiography for the abdominopelvic station and electrocardiogram-gated fast spin echo (ECG-FSE) MR angiography for the extremities. Two radiologists independently evaluated vessel segments for vascular stenosis, diagnosis confidence, graft presence, and Trans-Atlantic Inter-Society Consensus (TASC) II classification for each station. Diagnostic accuracies and κ agreement were assessed.

RESULTS

Of 573 vascular segments imaged, 16.9% (97/573, 19/20 patients) demonstrated hemodynamically significant abnormalities. Reader confidence was sufficient for diagnosis in 98% of segments with Gd MR angiography, 93% with QIR/ECG-FSE, and 95% with QISS. Overall reader confidence was higher with QISS than QIR/ECG-FSE within all 3 stations combined (P < .05). With low-confidence segments treated as misdiagnosis, sensitivity, specificity, positive predictive value, negative predictive value, accuracy, and κ agreement for all 3 stations combined were 81.4/87.2/57.0/95.8/86.2%/0.578 for QIR/ECG-FSE and 75.0/90.6/61.6/94.7/88.0%/0.597 for QISS. Using TASC II criteria to assess severity, QISS and QIR/ECG-FSE had no statistical difference in agreement with Gd MR angiography.

CONCLUSIONS

QISS and QIR/ECG-FSE MR angiography protocols demonstrate comparable diagnostic accuracies with high specificity. Either protocol provides an alternative to Gd MR angiography at 3T for patients with PAD.

Для цитирования: Hanrahan Christopher J, Lindley Marc D, Mueller Michelle, Kim Daniel, Sommers Daniel, Morrell Glen, Redd Andrew, Carlston Kristi, Lee Vivian S. Diagnostic Accuracy of Noncontrast MR Angiography Protocols at 3T for the Detection and Characterization of Lower Extremity Peripheral Arterial Disease.. Journal of vascular and interventional radiology : JVIR. 2018. DOI: 10.1016/j.jvir.2018.06.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.